– 61 –
Accessories
Main unit cover, mouthpiece, alkaline batteries, carrying
pouch, instruction manual. inhalation mask(S), inhalation
mask(L), AC adapter (optional)
The nebulizer gives off the high frequency sound and shuts off au
-
tomatically if the medication is not in contact with the mesh of the
medication chamber for more than 15 seconds (time varies for dif
-
ferent types of medication) or the medication is depleted. This is to
help prevent damage to the Mesh
TECHNICAL DATA:
Note:
Test result of cascade impact measurements for particle size.
• Performance may vary with different drugs such as suspensions
or high viscosity. See drugs supplier’s data sheet for further deta
-
ils.
• MMAD = Mass Median Aerodynamic Diameter.
Mass Median Aerodynamic Diameter.
MMAD 3.6μm
(Mass Median Aerodynamic Diameter
The Bottle Capacity
MAX.8ml
Noise
<50db(Range of 1m)
Particle size distribution compliant with EN 13544-1
individual tests
100
80
60
40
20
0
50
40
30
20
10
0
0,1
1
10
100
mean