Q155 Service Manual
1-3
cleaner applied with a damp cloth or sponge, then wipe dry.
n
The Salter Labs 7100 350c.c. Bubble Humidifier and the B&F Medical Bubble
Humidifier are suitable for use with the Oxygen Concentrators. Or a
humidifier capable of 30 psi operation and 15 LPM of flow is required. Use of
certain humidifiers not specified for use with this oxygen concentrator may
impair the performance.
n
Always choose the FDA or CE approved accessories, such as Masks or
cannulas & Humidifiers to ensure the reliability and performance.
n
Different personnel to prevent cross infection should never use the same mask
or cannula.
1.4.
Warranty
Merits Health Products Co., Ltd. warrants the Oxygen Concentrator to be free from
detect in general parts for a period of 5 years from the date of purchase excluding
the parts which warranty are specified on Table 1-1. Please tell your customer for
more details. This warranty is limited to the original purchaser of the Oxygen
Concentrator. Any defective part or assembly will be repaired or replaced, at the
sole discretion and determination of Merits Health Products Co., Ltd., if the unit has
been properly operated during the warranty period. Normal maintenance items and
disposable components (see Table 1-2) are not covered by this warranty. Shipping
charges, if any, shall be paid by the purchaser.
Note
: There is no other express warranty. Implied warranties, including those of
merchantability and fitness for a particular purpose, are limited to the
duration of the express limited warranty and to the extent permitted by law
and all implied warranties are excluded. This warranty does not cover
providing a loaner Oxygen Concentrator, compensating for costs incurred for
Oxygen Concentrator rental, or labor costs incurred in repairing or replacing
defective part(s).
Component
Part Number
Warranty Period
Note
Compressor
4-S01
10,000 hour
Molecular Sieve
2-S22
1 year
Table 1-1 List of specified warranty period components
Component
Part
Number
Note
Intake Foam
1-P19
Clean every week
Bacteria Filter
1-P43
Replace after every 10,000hr use
Intake Filter
3-P41
Replace on clean condition
Table 1-2 List of consumable