Q155 Service Manual
- 6-1 -
6.
Troubleshooting
SYMPTON
PROBABLE CAUSE
SOLUTION
No power on wall outlet.
Inspect house circuit breakers or fuses.
Tripped or faulty circuit
breaker.
Reset or replace the circuit breaker.
Faulty cable plug.
Replace the cable plug.
Power Loss while
power switch ON
Faulty power switch
Replace the power switch.
Faulty temperature sensor
Replace
the temperature
sensor
Air outlet block
Keep oxygen concentrator at least
three-inch away from walls, draperies
or furniture.
Faulty Cooling fan
Replace the cooling fan.
Over-Heating alarm
Environmental temperature
too high
Control room temperature below 35
℃
Faulty Solenoid Coil
Replace faulty Solenoid Coil
Clean foam filter or replace
Check the intake filter or replace it
Restriction in air passages
Check all the tubes from the exhaust of
air compressor to the oxygen outlet or
Replace the related faulty tubes
Faulty pressure sensor
Replace faulty pressure sensor
High Pressure alarm
Faulty Control Board
Replace Control Board
Faulty air compressor
Replace faulty air compressor
Low Pressure alarm
Leaking
Refer section 4.17 Leak Test and repair
leaking.
Leaking
Refer section 4.17 Leak Test and repair
leaking.
Faulty oxygen monitor
board
Replace faulty oxygen monitor board.
Faulty regulator
Replace faulty regulator or reset the
regulator
Faulty solenoid valve
Replace faulty solenoid valve
Relief valve release
Replace faulty relief valve
Faulty compressor flow
Replace or repair faulty compressor
Sieve bed contaminated or
molecule sieve invalid
Replace sieve bed
Environmental temperature
too high
Control room temperature below 35
℃
Restriction in exhaust
muffler
Replace or clean exhaust muffler
Clean foam filter or replace
Check the intake filter or replace it
Low Oxygen Purity
Restriction in air passages
Check all the tubes from the exhaust of
air compressor down to the oxygen
outlet or Replace the related faulty
tubes