The
illumipro-10
will complete the selected assay program and the Block Status on the MAIN MENU will
be shown as ’00:00’. Results for the selected assay can be viewed by selecting the Block displaying
completed status and following the instructions shown on the
illumipro-10
display. Results can be
printed manually by selecting the print option at the end of the results display or by enabling the Auto-
Print feature during System Set-up.
Upon completion of the assay run, the user must remove the
illumigene
devices from the unit and
discard appropriately.
NOTE:
Care must be taken to avoid contamination of the equipment and workspace by target
and/or amplified nucleic acids. DO NOT open illumigene devices after assay completion.
RESULTS MODE
The RESULTS MODE allows the user to view stored results, delete stored results and enable the auto-
printer. The
illumipro-10
will store up to 1000 individual test results or 200 Batches. The instrument
displays a warning to the user when result storage is approaching maximum capacity. RESULTS MODE
is accessed by following the instructions shown on the
illumipro-10
display.