21
APPENDIX
PRODUCT SPECIFICATIONS OVERVIEW
BASE UNIT:
10" L x 4.25" W x 1.65"D ABS plastic case with optional shock absorbing boot, 19ozs. including
batteries
DISPLAY:
128 x 128 pixel graphic display, 2.6” x 2.6” viewable area, backlight
POWER:
two (2) 9-volt alkaline batteries, two (2) 9-volt Lithium batteries, two (2) 9-volt NiMH batteries or optional
100 – 240 Vac power adapter (P/N A36742)
TEMPERATURE SPECIFICATIONS:
Storage: -40° F to 140° F (-40° C to 60° C)
Operating: 23° F to 122° F (-5° C to 50° C)
CONNECTIONS:
HART: Standard banana jacks (½” center)
PC Data transfer: DB-9 serial connector
Power: AC Adapter jack
AUTO SHUT-OFF
- occurs after user-selected period of keypad inactivity or can be disabled, both MFC and backlight
DIAGNOSTICS
– MFC 4100 self-tests upon power up. Advisories are posted on the display.
APPROVALS:
Class I, Div I, Groups A, B, C, D, T5 certified by MET Laboratories, Inc per UL3111-1, CAN/CSA-
22.2 No. 1010.1 & 157-92 and UL913, Fifth Edition, Rev. 2/21/97
CONTRAST
– direct adjustment of display contrast in any mode; automatic compensation for ambient temperature
changes
SECURITY LOCKOUT
– MFC 4100 Settings can be locked out entirely or by specific feature.
MEMORY
– separate, dedicated, non-volatile flash memory for MFC firmware, DOFs, and Configurations
DOF
Memory:
600 HART devices (based on average DOF size)
Configuration Memory:
200 HART devices (based on average configuration size)
ORDERING INFORMATION
BASE UNIT MODEL No & DESCRIPTION:
MFC 4100 HART Communicator
MFC 4101 HART Communicator with Documenting capability (via Meriam DMS software)
MFC 4102 HART Communicator with Intrinsically Safe certification. I.S. certification by MET
Laboratories per testing to UL3111-1 First Edition Rev. 6/30/94, CAN/CSA-22.2 No. 1010.1 &
157-92 and UL913 Fifth Edition
MFC 4103 HART Communicator with both Documenting and Intrinsically Safe certification
ORDER OPTIONS
-0 Rear hand strap only: no carrying case, no protective boot required
-1 Fitted, use-in-place, soft carrying case with adjustable shoulder strap and pocket for HART leads
-2 Slip-on, protective red boot with adjustable, detachable hanging strap
CE compliance