MTTR01 User Manual
AC Voltage
RANGE
RESOLUTION
ACCURACY
60mV
10uV
± (2.0% of rdg + 10D)
600mV
0.1mV
± (1.6% of rdg + 10D)
6V
1mV
± (1.5% of rdg + 10D)
60V
10mV
600V
0.1V
Input Impedance: >100MΩ for 60 and 600mV range, 10MΩ for 6, 60 and
600V.
Frequency Range: 40Hz ~ 400Hz
Overload Protection: 600Vrms
Reading: True rms
Max. Input voltage: 750Vac rms
DC Current
RANGE
RESOLUTION
ACCURACY
Remark
600uA
0.1uA
± (1.0% of rdg + 7D)
Auto range
6000uA
1uA
60mA
0.01mA
600mA
0.1mA
6A
1mA
± (1.5% of rdg + 7D)
10A
10mA
Overload Protection:
mA: F0.4A/600V fuse
10A: F10A/600V fuse (for measurement >5A, duration need to be less than
10 sec and use no more than once every 15 min)