Transducer Selection
The primary transducer (address 0) is selected by
default at power up. The
Local Operatio
n and
Remote Operation
sections of the manual show
how to select either transducer as the active
transducer through the LIMITS/SETUP menu in
manual operation, or with the _PCS4 XDUCER
command over a remote bus.
In addition to menu or bus selection, the F1 key
([2nd] [0]) is configured to toggle between the pri-
mary and the alternate internal transducers.
When commanding a transducer change, if the
current pressure is more than 10% (or 1 psi,
whichever is greater) over the maximum or under
the minimum range of the selected transducer the
change will not be accepted, and the display will
show the following message:
WARNING 01: UNABLE TO CHANGE TRANSDUCER
CURRENT READING OUT OF RANGE (USE CE)
Pneumatic Module
The pneumatic plumbing is arranged as shown on
the pneumatic schematic in figure 9.6. This module
includes a three-way solenoid valve (L6) installed
such that the MEASURE/CONTROL port is routed
to the primary transducer when the solenoid is
relaxed, or to the alternate transducer when it is
energized.
OPTIONS
PCS 400
8-6
www.mensor.com