Table 5.1 continued...
Continued on next page...
Symptoms and Clues
Possible Cause
Try This
4
SYSTEM DOES NOT CONTROL
AT COMMANDED SET POINT
CLUE: The PCS 400 requires an
input pressure higher than the
c o n t r o l o u t p u t s e t p o i n t , o r
pneumatics section is working
improperly.
The source pressure is not
high enough.
- - - - - - - - - - - - - - - - - - - - - -
The cable connecting the
e l e c t r i c a l m o d u l e t o t h e
regulator is disconnected.
- - - - - - - - - - - - - - - - - - - - - -
The solenoid valves are not
opening to let pressure into or
out of the control module.
- - - - - - - - - - - - - - - - - - - - - -
PCS 400 has a pneumatic
leak.
Apply source pressure 10 psi (or
10%) greater than the commanded
set point.
- - - - - - - - - - - - - - - - - - - - - - - - - -
Check the 9 pin connector
Å
on
the electrical module and the pcb
connector
Ã
o n t o p o f t h e
regulator. Make sure the regulator
connector is not shorted against
something.
- - - - - - - - - - - - - - - - - - - - - - - - - -
Run the solenoid valve test to verify
that all of the valves are working.
CAUTION: HIGH PRESSURES
COULD BE PRESENT AT THE
GAUGE OUTPUT DURING THE
TEST!
- - - - - - - - - - - - - - - - - - - - - - - - - -
See next section.
5
SYSTEM HAS A PNEUMATIC
LEAK
CLUE: Possible sources for leaks
are solenoids, plumbing, or relief
valves.
B e c a u s e o f t h e e x t r e m e
resolution of the PCS 400,
v e r y s m a l l t e m p e r a t u r e
changes on trapped volumes
of gas will appear as changes
in pressure. The PCS 400 will
control small leaks with no
e f f e c t
o n
a n y
c o n t r o l
parameters. The built-in leak
diagnostics are designed to
help troubleshoot problems
b u t n o t r e t u r n a b s o l u t e
numbers.
Run internal leak test and system
leak test to determine if leak is
inside unit or external. Most internal
leaks appear at solenoid valves,
relief valves and fittings.
6
SYSTEM DOES NOT RESPOND
F R O M T H E F R O N T P A N E L
KEYPAD
CLUE: Signals are not being
transferred from the front panel
keypad to the electrical module.
T h e
k e y p a d
c a b l e
i s
disconnected at the front
panel
È
or at the electrical
module.
Re-connect the cable.
MAINTENANCE
PCS 400
5-4
www.mensor.com
Summary of Contents for PCS 400
Page 13: ...User s Notes TABLE OF CONTENTS PCS 400 xii www mensor com ...
Page 23: ...User s Notes INSTALLATION PCS 400 2 4 www mensor com ...
Page 41: ...User s Notes LOCAL OPERATION PCS 400 3 18 www mensor com ...
Page 73: ...User s Notes SPECIFICATIONS PCS 400 7 4 www mensor com ...
Page 115: ...User s Notes INDEX PCS 400 10 6 www mensor com ...
Page 116: ......