Calibration Software PTC04
User Manual
Page 10 of 17
REVISION 1.44.01 - 2 MAY 2018
UI_0PTC04AAMLX
2.
The program notifies you that the current calibration is within tolerance! No Correction of the calibration
parameters (Calibration) is needed.
Figure 16: Calibrate Anyway?
If no calibration is needed you can still choose to calibrate or not.
If you press yes, the new calibration parameters are written into the PTC04.
If you press no, there are no modifications done.
3.
At the end of the calibration, the software verifies if the programmer was at temperature when the calibration
started. If the programmer was not at temperature, the following message is given.
Figure 17: Programmer is not at temperature.
It is advised not to calibrate (store the new calibration parameters in the programmer) when the programmer is not
at operating temperature. This is important for the accuracy of the PTC04 programmer.
If you press
yes
, the new calibration parameters are written into the PTC04.
If you press
no
, there are no modifications done.