Calibration Software PTC04
User Manual
Page 9 of 17
REVISION 1.44.01 - 2 MAY 2018
UI_0PTC04AAMLX
The calibration check
Now the calibration software will perform a number of measurements on all the ADC and DAC channels of the
programmer. This can take several minutes. At the end of the calibration check the program verifies if a correction of
the calibration parameters is needed.
Figure 14: Calibration Check
There are three possible outcomes
1.
The program notifies you that the current calibration is out of tolerance and ask you if you want to corrected the
calibration parameters or not.
Figure 15: Calibrate?
If you press
yes
, the new calibration parameters are written into the PTC04.
If you press
no
, there are no modifications done.