background image

Chapter 5 – Logging  

13 

Chapter 5 – Logging 

 

DIN EN ISO 11607-2 requires that the sealing process be monitored and documented in order to satisfy 
the requirement for compliance with the specifications contained in part 2 of the standard. 
As a result, you are able to record the logs of the seal runs and output and archive them on the following 
output media: 

 

Log printer MELAprint 42/44 

 

MELAflash CF-Card printer (from version 1.9) 

 

Documentation software MELAview (from version 2.2)/MELAtrace 

Log printer as output medium  

If you want to employ the log printer MELAprint 42/44 as the output medium, connect it to the sealing 
device as follows: 

1. 

Connect the cables of the log printer to the printer connection (RS232 interface) on the back side 
of the sealing device. 

2. 

Connect the power supply cable included in the scope of delivery of the log printer in accordance 
with the appropriate operating instructions. 

The sealing device recognizes the log printer automatically. 
The log printer issues a log row for every sealing procedure performed. 

MELAflash CF-Card Printer as output medium 

The MELAflash CF-Card printer serves for storage of logs on the MELAflash CF card. The readout of the 
logs from the CF card to the practice computer is carried out with the MELAflash card reader. A CF card 
and the card reader are included in the scope of delivery of the MELAflash CF-Card printer. 
MELAflash CF-Card-Printer (from software version 1.9) supports the recording of logs for the MELAseal 
Pro rotary sealer. 

If you want to employ the MELAflash CF-Card printer as the output medium, connect it to the autoclave as 
follows: 

1. 

Connect the cables of the log printer to the printer connection (RS232 interface) on the back side 
of the sealing device. 

2. 

Connect the power supply cable included in the scope of delivery of the log printer in accordance 
with the appropriate operating instructions. 

The sealing device recognizes the MELAflash CF-Card printer automatically. A daily log is saved on the 
CF card in the MELAflash CF-Card Printer after the first sealing procedure of the day. Here a log row is 
written for every further sealing procedure of the same day. 

 

PLEASE NOTE 

To ensure complete documentation, ensure that the CF card is inserted in the MELAflash CF-
Card Printer during every sealing procedure. 

 

Using the computer as an output medium 

In order to be able to use a computer as an output medium, the computer must be connected via the 
series interface on the rear of the sealing device. 
Please use the documentation software MELAview (from version 2.2)/MELAtrace to read out the logs. 
How to run and read out logs via the program MELAview/MELAtrac

e

 is outlined in the internal help 

function in the program. 
 
The computer generates daily logs in the same manner as the MELAflash

 

CF-Card printer; i.e. a daily log 

is entered in the computer following the first sealing procedure of the day. Here a log row is written for 
every further sealing procedure of the same day. 

 

 

Summary of Contents for MELAseal Pro

Page 1: ...reatment and hygiene field You our customer are justified in your demand for the best products quality and reliability Providing competence in hygiene and Quality made in Germany we guarantee that the...

Page 2: ...a component of the product About this manual Symbols used Symbol Explanation Indicates a dangerous situation which if not avoided could entail slight to life threatening injuries Indicates a dangerou...

Page 3: ...t medium 13 MELAflash CF Card Printer as output medium 13 Using the computer as an output medium 13 Opening and printing log files on the computer 14 Log folder structure 14 File name and ending 14 Re...

Page 4: ...other than a member of a MELAG authorized technical customer service Fan The fan in the base of the sealing device must be free at all times and may not be allowed to become blocked or congested Failu...

Page 5: ...al Device Directive Suitable materials It is suitable for the heat sealing of the following packaging materials Heat sealable transparent sterilization packaging e g MELAfol Should you wish to use pac...

Page 6: ...seal Pro rotary sealer Power cable Operating Manual Views of the device Fig 1 Views of the device 1 Guide rail for straight introduction of the sterilization packaging 2 Operating and display field 3...

Page 7: ...increase or reduce the sealing temperature Reverse key for removing the sterilization packaging upon a blockage of the conveyor belts Daily batch counter for increasing or reducing the total number o...

Page 8: ...hould be located at least 1 5 m away radius from the treatment area Observe all the information contained in this chapter during commissioning DANGER Failure to comply with the set up conditions can r...

Page 9: ...aling device 1 Loosen the screw a little in order to slide the guide rail forwards or backwards The scale to the right of the adjusting screw displays the overhang in mm measured from the outer edge o...

Page 10: ...is could result in the deposit of film residue on the sealing rail Using the reverse key Should either the sterilization packaging become jammed in the device or the conveyor belt come to a halt you c...

Page 11: ...the current date proceed as follows 1 Activate the sealing device at the power switch 2 Press as soon as the display shows the current software version 3 Press repeatedly to access the Date menu 4 Pr...

Page 12: ...the Signal tone It is possible to set a signal tone to control the number of packages to be sealed It will sound after the pre set number of seals has been completed This function is useful when need...

Page 13: ...ELAflash CF Card Printer from software version 1 9 supports the recording of logs for the MELAseal Pro rotary sealer If you want to employ the MELAflash CF Card printer as the output medium connect it...

Page 14: ...Card and the computer The directory name consists of five characters identical with the first five characters of every log e g 9K0EQ Under this folder is a sub directory containing the month of the pr...

Page 15: ...ace Example of a log row for a successfully concluded sealing procedure 01161 2009SPF1000 09 12 15 14 28 180 099 88 185 175 120 090 120 078 F300 Key The following parameters are displayed successively...

Page 16: ...actical and inexpensive solution suitable for sealing light packaging Roll dispenser The roll dispenser represents the ideal assistant when you need to process MELAfol transparent sterilization packag...

Page 17: ...t suitable for use with medical products It should not deposit any oily residue After 1 year Clean the fan on the base of the sealing device once a year in order to prevent it from becoming clogged wi...

Page 18: ...rned The sealing temperature has not been set correctly Check the sealing temperatures set on the sealing device and if necessary change the setting see page 12 On repeated occurrence inform your MELA...

Page 19: ...customer services Temp too high The sealing temperature is too high for a short period If the malfunction message disappears after a few minutes you can continue operation as usual On repeated occurr...

Page 20: ...ion documentation within the scope of process clearance Check for dryness and integrity once again before using the instruments Weekly Checking the seal seam with aids Perform a visual control using M...

Page 21: ...st price of 125 00 plus VAT state 01 2017 After the film test strips have been tested MELAG will issue verification of testing and if successful a certificate This certificate confirms conformity of t...

Page 22: ...omposite film should be filled with a max of up to 3 4 of its volume DIN 58953 7 2010 When using transparent sterilization packaging from a roll the removal side must have an overlap of min 1 cm betwe...

Page 23: ...thus not be generally specified Table 2 only makes recommendations regarding the storage length of sterile medical products The following requirements apply to the storage of sterile medical products...

Page 24: ...dapter 80270 Replacement parts Separating plate with roll dispenser 71200 Angle adapter for RS232 socket 80210 Device fuses 3 15 A T 5x20 74320 Technical Data Model name MELAseal Pro Device dimensions...

Page 25: ......

Page 26: ...a e 6 10 10829 Berlin Germany E Mail info melag de Web www melag com Responsible for content Technical Office We reserve the right to technical alterations Your stockist BA_sealPro_EN_v6 docx Rev 6 17...

Reviews: