Chapter 3 – Commissioning
10
As long as the sealing temperature has not been reached, the display shows the message
Heating up
.
Minimal temperature differences of ± 5 °C are possible during the sealing procedure.
In the case of deviations over 5 °C, wait until the sealing device display shows Status:
OK
.
Sealing procedure
As soon as the sealing temperature has been reached, you can insert the sterilization packaging as
follows:
1.
Insert the sterilization packaging in the guide on the left-hand side of the device.
The conveyor belt begins to turn as soon as the inbuilt optical sensor recognizes the packaging. The
packaging is then guided through the sealing device automatically.
WARNING
The film side of the packaging must always face upwards. Should the packaging be inserted
incorrectly, this could result in the deposit of film residue on the sealing rail.
Using the reverse key
Should either the sterilization packaging become jammed in the device or the conveyor belt come to a halt,
you can reverse the conveyor belt by pressing the reverse key
. This enables the operator to remove
the packaging on the left-hand side of the device.
■ If it is not possible to remove the packaging by pressing the reverse key, please consult your
stockist or the MELAG customer service department.
WARNING!
Failure to observe this provision could result in damage to or malfunction of the sealing
device.
Do not attempt to remove the packaging from the device with violence.
Display of the daily batches
The daily batch counter is displayed in the upper corner of the display (#: 000). It counts the number of
seals carried out in the course of one day. If the sealing device is switched off, the counter is reset to #000.
PLEASE NOTE
This does not reset the total batch counter.
Temp:086 °C #:000
Heating up
Summary of Contents for MELAseal Pro
Page 25: ......