MELAG MELAdest 65 Operating Manual Download Page 7

Chapter 2 – Commissioning 

Before commissioning 

Cleaning the distilling chamber 

Although the distilling chamber of the MELA

dest

 65 has been cleaned with the greatest of care, 

manufacturing, packaging and dispatch procedures could result in the presence of harmless residues of 

various substances. As a result, the distilled water produced during initial commissioning could assume an 

undesired odour. To eliminate all odour-generating factors, proceed as follows: 

1. 

Setup the water distiller.  

2. 

Remove the connection cable from the distilling chamber. 

3. 

Remove the upper section of the water distilling chamber and fill with c. 0.5 L tap water. 

4. 

Return the upper section to the distilling chamber and connect the connection cable to the 

bushing on the side wall of the chamber. 

5. 

Replace the glass flask laterally rotated without the adapter cover under the distilling chamber 

discharge opening and plug the plug into the socket.  

                                   

 

6. 

Press the START/RESET button to activate the water distiller.  

The ventilator motor begins to run. 

The water distiller deactivates automatically after the water has boiled (no steam egress from the 

upper section discharge opening.  

7. 

Remove the mains plug and empty the glass flask.  

The distilling chamber is now flushed and ready for commissioning. 

 

Checking the cover seal in the upper section of the distilling chamber  

Before initial commissioning and every further use of the device, check the correct position of the cover 

seal in the upper section of the distilling chamber. 
The seal may not present any uneven surfaces and must lay entirely flat so that it maintains the seal. 
Otherwise, boiling water could leave the upper section and e.g. flow onto the work surface. 
If the seal does not lay flat, press with the hand. Do not use any sharp objects. 
 

 

incorrect position of the seal

 

sealing lip

 

sealing lip

 

seal evenly fitted at all points

 

 

Discharge opening 

of the distilling 

chamber 

Summary of Contents for MELAdest 65

Page 1: ...ea of practice hygiene since 1951 During this time we advanced to become one of the leading manufacturers of sterilizers World sales of more than 450 000 MELAG devices gives testimony to the quality o...

Page 2: ...e device It represents a component of the product User group These instructions are addressed to doctors their assistants and service departments About this manual Symbols used Symbol Explanation Indi...

Page 3: ...e 4 Scope of delivery 4 Views of the device 4 Chapter 2 Commissioning 5 Fitting the glass flask 5 Before commissioning 7 Chapter 3 Operation and care 8 Producing distilled water 8 Cleaning the distill...

Page 4: ...o 100 C and begins to boil During evaporation the chemicals salts and soiling remain in the container The stream flows through the cooling coil system in the upper section where it cools and condenses...

Page 5: ...ollar on the neck of the glass flask in such a way that the other screws on the collar lie exactly opposite the spout cap pos a 6 Refasten the metal collar with both screws PLEASE NOTE Do not overtigh...

Page 6: ...11 Insert the handle cover pos e and the small hook pos f on the end of the handle end so that they snap closed A bracket for the adapter cover is integrated in the handle cover To this end the adapt...

Page 7: ...distilling chamber discharge opening and plug the plug into the socket 6 Press the START RESET button to activate the water distiller The ventilator motor begins to run The water distiller deactivate...

Page 8: ...e of hot water reduces the length of the process 4 Replace the upper section on the distilling chamber and reconnect the connection cable to the distilling chamber bushing PLEASE NOTE Check the seal i...

Page 9: ...stilling chamber with lukewarm tap water and a sponge to remove the deposits from the distilling chamber With stubborn soiling deposits on the inside walls use only a mild stainless steel cleaning med...

Page 10: ...ciently water escapes Glass flask positioned incorrectly under the water distiller Cooling system soiled Connect the connection cable to the distilling chamber Refill with tap water Replace the cover...

Page 11: ...er section 65410 MELAdest 65 housing upper section 65420 MELAdest 65 cover seal for housing upper section 65510 Technical Data Model name MELAdest 65 Exterior dimensions of the water distiller x H 23...

Page 12: ...ntechnik oHG Geneststra e 6 10 10829 Berlin Germany E Mail info melag de Web www melag de Responsible for content Technical Office Subject to technical alterations Your stockist BA_EN_dest65_v2 docx R...

Reviews: