Chapter 4 – Malfunctions 

10 

Chapter 4 – Malfunctions 

 

Incident 

Possible cause 

What you can do 

No distilled water is produced 

The mains plug is not in the 

mains socket 

 

Mains socket is defective 

 

 

Upper section connector cable is 

not on the distilling chamber 

 

START/RESET button is not 

depressed 

Check the mains plug 

 

 

Check the electricity supply on 

the mains socket 

 

Connect the connection cable to 

the distilling chamber 

 

Press the START/RESET button 

Too little distilled water produced  Upper section connection cable is 

not in the distilling chamber 

 

Distilling chamber is not filled 

sufficiently 

 

The cover seal does not seal 

sufficiently, water escapes 

 

 

Glass flask positioned incorrectly 

under the water distiller 

 

 

 

Cooling system soiled 

Connect the connection cable to 

the distilling chamber 

 

Refill with tap water 

 

 

Replace the cover seal in the 

upper section of the distilling 

chamber  

 

Check whether water flows past 

the glass flask and place the 

glass flask under the water 

distiller correctly 

 

Clean the cooling system, blow 

out with compressed air if 

necessary 

Steam escapes under the 

discharge opening of the upper 

section 

Upper section connection cable is 

not in the distilling chamber 

 

Connect the connection cable to 

the distilling chamber 

 

The water distiller does not heat  The water distiller overheats 

Allow the water distiller to cool for 

5-10 minutes, then restart 

The water distiller does not 

deactivate automatically 

Overheating switch defective 

 

 

Summary of Contents for MELAdest 65

Page 1: ...ea of practice hygiene since 1951 During this time we advanced to become one of the leading manufacturers of sterilizers World sales of more than 450 000 MELAG devices gives testimony to the quality o...

Page 2: ...e device It represents a component of the product User group These instructions are addressed to doctors their assistants and service departments About this manual Symbols used Symbol Explanation Indi...

Page 3: ...e 4 Scope of delivery 4 Views of the device 4 Chapter 2 Commissioning 5 Fitting the glass flask 5 Before commissioning 7 Chapter 3 Operation and care 8 Producing distilled water 8 Cleaning the distill...

Page 4: ...o 100 C and begins to boil During evaporation the chemicals salts and soiling remain in the container The stream flows through the cooling coil system in the upper section where it cools and condenses...

Page 5: ...ollar on the neck of the glass flask in such a way that the other screws on the collar lie exactly opposite the spout cap pos a 6 Refasten the metal collar with both screws PLEASE NOTE Do not overtigh...

Page 6: ...11 Insert the handle cover pos e and the small hook pos f on the end of the handle end so that they snap closed A bracket for the adapter cover is integrated in the handle cover To this end the adapt...

Page 7: ...distilling chamber discharge opening and plug the plug into the socket 6 Press the START RESET button to activate the water distiller The ventilator motor begins to run The water distiller deactivate...

Page 8: ...e of hot water reduces the length of the process 4 Replace the upper section on the distilling chamber and reconnect the connection cable to the distilling chamber bushing PLEASE NOTE Check the seal i...

Page 9: ...stilling chamber with lukewarm tap water and a sponge to remove the deposits from the distilling chamber With stubborn soiling deposits on the inside walls use only a mild stainless steel cleaning med...

Page 10: ...ciently water escapes Glass flask positioned incorrectly under the water distiller Cooling system soiled Connect the connection cable to the distilling chamber Refill with tap water Replace the cover...

Page 11: ...er section 65410 MELAdest 65 housing upper section 65420 MELAdest 65 cover seal for housing upper section 65510 Technical Data Model name MELAdest 65 Exterior dimensions of the water distiller x H 23...

Page 12: ...ntechnik oHG Geneststra e 6 10 10829 Berlin Germany E Mail info melag de Web www melag de Responsible for content Technical Office Subject to technical alterations Your stockist BA_EN_dest65_v2 docx R...

Reviews: