General notes 

Please read this operating manual carefully before commissioning the device. The instructions include important safety 

information. The functionality and value-retention of this autoclave depends on the care accorded to it. 
Please store this operating manual carefully and in close proximity to the device. It represents a component of the 

product. 

User group 

These instructions are addressed to doctors, their assistants and service departments. 

About this manual

 

Symbols used 

Symbol 

Explanation 

 

Indicates a dangerous situation, which if not avoided could result in injuries ranging in 

seriousness from light to life-threatening. 

 

Draws your attention to a situation, which if not avoided, could result in damage to the 

instruments, the practice fittings or the device. 

 

Draws your attention to important information. 

 

Summary of Contents for MELAdest 65

Page 1: ...ea of practice hygiene since 1951 During this time we advanced to become one of the leading manufacturers of sterilizers World sales of more than 450 000 MELAG devices gives testimony to the quality o...

Page 2: ...e device It represents a component of the product User group These instructions are addressed to doctors their assistants and service departments About this manual Symbols used Symbol Explanation Indi...

Page 3: ...e 4 Scope of delivery 4 Views of the device 4 Chapter 2 Commissioning 5 Fitting the glass flask 5 Before commissioning 7 Chapter 3 Operation and care 8 Producing distilled water 8 Cleaning the distill...

Page 4: ...o 100 C and begins to boil During evaporation the chemicals salts and soiling remain in the container The stream flows through the cooling coil system in the upper section where it cools and condenses...

Page 5: ...ollar on the neck of the glass flask in such a way that the other screws on the collar lie exactly opposite the spout cap pos a 6 Refasten the metal collar with both screws PLEASE NOTE Do not overtigh...

Page 6: ...11 Insert the handle cover pos e and the small hook pos f on the end of the handle end so that they snap closed A bracket for the adapter cover is integrated in the handle cover To this end the adapt...

Page 7: ...distilling chamber discharge opening and plug the plug into the socket 6 Press the START RESET button to activate the water distiller The ventilator motor begins to run The water distiller deactivate...

Page 8: ...e of hot water reduces the length of the process 4 Replace the upper section on the distilling chamber and reconnect the connection cable to the distilling chamber bushing PLEASE NOTE Check the seal i...

Page 9: ...stilling chamber with lukewarm tap water and a sponge to remove the deposits from the distilling chamber With stubborn soiling deposits on the inside walls use only a mild stainless steel cleaning med...

Page 10: ...ciently water escapes Glass flask positioned incorrectly under the water distiller Cooling system soiled Connect the connection cable to the distilling chamber Refill with tap water Replace the cover...

Page 11: ...er section 65410 MELAdest 65 housing upper section 65420 MELAdest 65 cover seal for housing upper section 65510 Technical Data Model name MELAdest 65 Exterior dimensions of the water distiller x H 23...

Page 12: ...ntechnik oHG Geneststra e 6 10 10829 Berlin Germany E Mail info melag de Web www melag de Responsible for content Technical Office Subject to technical alterations Your stockist BA_EN_dest65_v2 docx R...