1.4.7 Data-Display
The module „Data-Display“ displays saved data. How this data is
displayed depends on the type of data. Analog signals are displayed
in a window that resembles the oscilloscope mode. Digital signals
are displayed in a window that resembles the Logic Analyzer mode.
1.4.8 Set up Offline Mode
The MEphisto Scope UM203 may save data onto an SD-card that
has a capacity of up to 2 GB. After applying the supply voltage, the
MEphisto Scope uses the settings that were last made online. In-
serting the SD-card starts the measurement process.
Notes: Whether offline or online via PC, the way how the MEphisto
Scope (UM203) measures is exactly the same. However, there are
some aspects to consider.
The most important difference between online and offline mode is
the reference potential. Within the online-mode, the device is con-
nected to the mass of the PC. If it is a Desktop-PC, the device’s