10 

5. Troubleshooting 

                                                               

SZG series

 

                     

The performance of the microscope can’t be made fully because of unfamiliar using, this table will give some 

advices: 

 

Trouble Cause 

Remedy 

1. Optical parts 

 

Brightness too bright or 

too dark 

Adjustment of brightness is not 
correctly 

Adjust it correctly 

Dirt on specimen 

Clean specimen 

Dirt on surface of eyepiece 

Clean eyepiece 

Dirt on surface of objective 

Clean objective 

Dirt appears inside of 

view field 

Dirt on stage 

Clean stage 

Interpupillary is not right 

Re-adjust the interpupillary 

Diopter is not right 

Re-adjust the diopter 

 Double image 

Different magnification of both 
eyepiece 

Use the same eyepiece 

 Image is not clear 

Dirt on surface of objective 

Clean objective 

Diopter is not right 

Re-adjust the diopter 

 Image blur during 

focusing 

Wrong focusing 

Focus again 

 Incision image appears 

when observe through 
eyepiece or TV device 

Light path selection lever is not 
at right position 

Turn it to right position 

 Image is not clear on 

monitor during focusing 

Wrong focus depth of video 
device 

Re-adjust the focus depth by adjusting 
ring on TV adapter 

2. Electrical parts 

Wrong setting of bulb 

Set the bulb correctly 

Bulb has burned out 

Replace the bulb 

Fuse has burned out 

Replace the fuse 

 Bulb does not work 

Dirt on the stage stopped the 
light path 

Clean the stage 

 Bulb burned out   

The bulb is not standard 

Use standard bulb 

Bulb is going to burn out 

Replace the bulb 

 

Light flickered out 

The wires do not connect right 

Connect it correctly 

3. Focusing parts 

 Focusing knobs work 

stiff 

Too tight of the focusing knobs 

Loose it to suitable 

 Image is not clear by 

the focusing arm slipping 
down during focusing 

To loose of the focusing knobs 

Tight it to suitable 

 

Summary of Contents for GMT-1 BF/DF

Page 1: ...GMZT PT 76 GMZT PT 85 GMZ 5 BF BD GMZ 5TR BF BD GMZ 8TR BF BD GMZ 13 BF BD GMZ 13TR BF BD GMZ 40 BF DF GMZ 41 BF DF GMZ 50 BF DF GMZ 51 BF DF GMT 1 BF DF GMT 2 BF DF GMT 3 BF DF...

Page 2: ...erpupillary distance 5 3 6 Auxiliary lens 5 3 7 Mount and remove micrometer 6 3 8 Set photo device TV device 6 3 9 Select light path 6 3 10 Focus TV device 6 3 11 Iris diaphragm 7 3 12 Switch Bright f...

Page 3: ...ise it will with trouble 6 Please shut off power supply before replace bulb 7 Please shut off power supply before remove the microscope 8 Please holder the microscope from arm and base and put off saf...

Page 4: ...1 1 Nomenclature SZG series Microscope head Lock screw Eyepiece Focusing knob Microscope stand Gem clamp...

Page 5: ...le Switch for top light Zoom knob Screw to hold top light Switch for iris diaphragm Knob for darkfield brightfield Plug Lock screw for light holder Screw for rotatable axial Brightness adjusting knob...

Page 6: ...SZN series the numbers show assemblage order Before assemblage please make sure that every part is clean And keep them clean during assemblage 135 camera C type monitor PK mount Photo eyepiece Photo t...

Page 7: ...se Hold the front of the base and arm to turn to suitable position for observation Fig 2 Then tight the screw for rotatable axial The plastic disc under base can make the whole microscope for 360 rota...

Page 8: ...ob to Min magnification then observe through left tube If the image is not clear adjust diopter to make it clear Fig 4 3 5 Interpupillary distance Hold both eyepiece tube to turn as the direction show...

Page 9: ...ew 3 9 Select light path To observe through binocular please turn the light path selection lever to OUT Fig 10 If turn the light path selection lever to IN Fig 10 you can observe through both binocula...

Page 10: ...ight It is 7W fluorescent bulb for top light and connect by a soft tube Fig 12 3 14 Replace fuse Please set the power switch at O before replace fuse Use screw driver to take off fuse from its base an...

Page 11: ...on cover and turn over it Pull out the bulb for replacement Fig 15 Please wait the bulb cool down for replacement Don t touch the bulb directly by hand 3 17 Polarizer There are 2 piece of polarizer O...

Page 12: ...45 0X 5 3 60 0X 4 0 22 5X 10 7 45 0X 5 3 67 5X 3 6 90 0X 2 7 33 8X 7 1 60 0X 4 0 90 0X 2 7 120 0X 2 0 45 0X 5 3 15X 16 67 5X 3 6 101 3X 2 4 135 0X 1 8 50 6X 4 7 13 4X 17 9 20 1X 11 9 26 8X 9 0 10 1X 2...

Page 13: ...r Image blur during focusing Wrong focusing Focus again Incision image appears when observe through eyepiece or TV device Light path selection lever is not at right position Turn it to right position...

Reviews: