background image

                                                                                             

SZG series   

 3-4   Diopter & focusing                                                     

                                                                       

 Turn the focusing knob to Max. magnification. 

                                                                       

 Turn the both eyepiece tube at “0” position. 

                                                                       

 Observe through right tube. If the image is not clear, use 

focusing knob to make it clear. 

                                                                       

 Turn the focusing knob to Min. magnification. 

                                                                       

 Observe through right tube. If the image is not clear, 

adjust diopter to make it clear.(Fig. 4) 

                                                                       

 Turn the focusing knob to Max. magnification again. If 

the image is still not clear, repeat the above step 

 and 

                                                                       

 Turn the focusing knob to Min. magnification, then 

observe through left tube. If the image is not clear, 

adjust diopter to make it clear. (Fig. 4) 

                                                                               

             

 

 3-5   Interpupillary distance                                                       

                                                                                        Hold both eyepiece tube to turn as the direction showed   

Fig. 5 to suitable position 

                                                                   

 

 

                                                                               

 3-6   Auxiliary lens 

                                                                       

 Set on auxiliary lens to the bottom of the microscope head. 

                                                                     

 ⑵

 It should use additional tube when use 0.5X auxiliary 

lens. (Fig. 6) 

                                                                               

 When use 1.5X or 2X auxiliary lens, please take off 4pcs   

M3 screws on focus arm and set the focus arm into 4pcs 

holes and lock. (Fig. 7) 

                                                                         

 It need not use additional tube when use 0.75X 

auxiliary lens. 

                                                                                   

 

 

 

 

 

 

Figure 4 

 

Figure 5 

 

Figure 6 

Left diopter 

Microscope 

head

 

Additional tube 

Focus arm 

 

Figure 7 

Right diopter 

Summary of Contents for GMT-1 BF/DF

Page 1: ...GMZT PT 76 GMZT PT 85 GMZ 5 BF BD GMZ 5TR BF BD GMZ 8TR BF BD GMZ 13 BF BD GMZ 13TR BF BD GMZ 40 BF DF GMZ 41 BF DF GMZ 50 BF DF GMZ 51 BF DF GMT 1 BF DF GMT 2 BF DF GMT 3 BF DF...

Page 2: ...erpupillary distance 5 3 6 Auxiliary lens 5 3 7 Mount and remove micrometer 6 3 8 Set photo device TV device 6 3 9 Select light path 6 3 10 Focus TV device 6 3 11 Iris diaphragm 7 3 12 Switch Bright f...

Page 3: ...ise it will with trouble 6 Please shut off power supply before replace bulb 7 Please shut off power supply before remove the microscope 8 Please holder the microscope from arm and base and put off saf...

Page 4: ...1 1 Nomenclature SZG series Microscope head Lock screw Eyepiece Focusing knob Microscope stand Gem clamp...

Page 5: ...le Switch for top light Zoom knob Screw to hold top light Switch for iris diaphragm Knob for darkfield brightfield Plug Lock screw for light holder Screw for rotatable axial Brightness adjusting knob...

Page 6: ...SZN series the numbers show assemblage order Before assemblage please make sure that every part is clean And keep them clean during assemblage 135 camera C type monitor PK mount Photo eyepiece Photo t...

Page 7: ...se Hold the front of the base and arm to turn to suitable position for observation Fig 2 Then tight the screw for rotatable axial The plastic disc under base can make the whole microscope for 360 rota...

Page 8: ...ob to Min magnification then observe through left tube If the image is not clear adjust diopter to make it clear Fig 4 3 5 Interpupillary distance Hold both eyepiece tube to turn as the direction show...

Page 9: ...ew 3 9 Select light path To observe through binocular please turn the light path selection lever to OUT Fig 10 If turn the light path selection lever to IN Fig 10 you can observe through both binocula...

Page 10: ...ight It is 7W fluorescent bulb for top light and connect by a soft tube Fig 12 3 14 Replace fuse Please set the power switch at O before replace fuse Use screw driver to take off fuse from its base an...

Page 11: ...on cover and turn over it Pull out the bulb for replacement Fig 15 Please wait the bulb cool down for replacement Don t touch the bulb directly by hand 3 17 Polarizer There are 2 piece of polarizer O...

Page 12: ...45 0X 5 3 60 0X 4 0 22 5X 10 7 45 0X 5 3 67 5X 3 6 90 0X 2 7 33 8X 7 1 60 0X 4 0 90 0X 2 7 120 0X 2 0 45 0X 5 3 15X 16 67 5X 3 6 101 3X 2 4 135 0X 1 8 50 6X 4 7 13 4X 17 9 20 1X 11 9 26 8X 9 0 10 1X 2...

Page 13: ...r Image blur during focusing Wrong focusing Focus again Incision image appears when observe through eyepiece or TV device Light path selection lever is not at right position Turn it to right position...

Reviews: