Press the “BACK” button to leave the screen.
B. TM1800 as reference for calibrating a
resistive transducer
1]
Press the “Calibrate” button. The following
question appears.
2]
Press the “YES” button. The following ques-
tion appears.
3]
Press the “NO” button. The following mes-
sage appears.
4]
Connect the transducer to the specified chan-
nel. Confirm by pressing “OK”. The following
screen appears.
5]
Move the transducer sliding contact to the
desired position and keep it there. Enter the
“Distance (mm)” value in the field “First
point”, see screen above. The value in the
corresponding “Voltage (mV)” field will fol-
low automatically.
6]
Move the cursor to the entry field for the
second point. For quick navigation between
the fields use the arrow buttons next to it.
Proceed in the same way to calibrate the
remaining positions. The calibration data of
each position is stored automatically when
moving the cursor to the next field.
7]
Type in a significant name for the transducer
into the “Transducer ID” field.
8]
Press the “Finish Calibration” button.
9]
Press the “BACK” button to leave this screen.
Summary of Contents for TM1800
Page 2: ......
Page 7: ...CG0757JE ZP CG01E TM1800 7...
Page 17: ...CG0757JE ZP CG01E TM1800 17 3 System overview...
Page 25: ...CG0757JE ZP CG01E TM1800 25 3 System overview Keyboard in position 3...
Page 71: ...CG0757JE ZP CG01E TM1800 71 7 Perform testing...
Page 97: ...CG0757JE ZP CG01E TM1800 97 12 Quick Guide...
Page 103: ......