Maintenance and care
29
5
Maintenance and care
Repair and maintenance work has to be carried out by Megger or authorised service
partners using original spare parts only. Megger recommends having the system tested
and maintained at a Megger service centre once a year.
Megger also offers its customers on-site service. Please contact your service centre if
needed.
It is not necessary to open the housing of the device to commission and operate the
system. Opening the housing causes the immediate termination of all warranty claims!
The connections and connection leads of the system must be regularly tested to ensure
that they are free of defects and intact, in accordance with the applicable national and
company-specific arrangements.
If the device is not used for a lengthy period, it should be stored in a dust-free and dry
environment.
Do not clean the display with aggressive products such as solvents or spirits.
Instead, use lukewarm water and a soft, lint-free cloth for wet wiping, or a microfibre
cloth for dry wiping.
Each connection cable is fused in the area of the plug connector with a T 25A H 440V
micro-fuse (6.3 mm x 32 mm) which may be replaced independently if need be.
Repair and
maintenance
Storage
Caring for the display
Replacing fuses