Figure 3.
Wrap excess extensions around the perimeter of the neurostimulator.
2.
Use the suture holes in the connector block to secure the neurostimulator to the
muscle fascia with nonabsorbable silk.
Checking system integrity
Caution:
To use the nonsterile clinician programmer in a sterile field, place a sterile
barrier between the patient and the programming head to prevent infection. Do not
sterilize any part of the clinician programmer. Sterilization may damage the
programmer.
Note:
The neurostimulator should be in the pocket during system interrogation for integrity
to ensure proper readings.
1.
To ensure proper connection of each extension to the neurostimulator, use the
clinician programmer to program the basic stimulation parameters, check the battery
status, and check the electrode impedances to rule out a short or open circuit.
2.
If the system integrity test results are not acceptable, refer to "Connecting the
extension to the neurostimulator" on page 12.
Completing the implant procedure
1.
Close and dress all incisions.
2.
Ensure that a patient control device is given to the patient.
3.
Complete the device tracking and patient registration paperwork and return the
documents to Medtronic.