Socket II (Electrodes 8-15)
Socket I (Electrodes 0-7)
Extension 2
Extension 1
Figure 1.
Insert the extension connectors into the neurostimulator.
Note:
Insert a connector plug (from an accessory kit) into unused neurostimulator
socket.
4.
For each extension or plug, fully insert the torque wrench (packaged with the
neurostimulator) into each self-sealing grommet of the connector block and tighten
each setscrew (Figure 2).
Cautions:
▪
To prevent undertightening the neurostimulator setscrews, do not use the
torque wrench from the extension kit. Undertightening may result in
insufficient electrical contact within the connector block, which may cause
intermittent stimulation.
▪
Be sure the torque wrench is fully inserted into the self-sealing grommet. If
the torque wrench is not fully inserted, the setscrew may be damaged,
resulting in intermittent or loss of stimulation.
▪
Before tightening setscrews, ensure that the extension connector is inserted
into the connector block to prevent damaging the extension.
▪
Verify that each leaf of the self-sealing grommet is closed after the torque
wrench is withdrawn. If fluid leaks through a grommet seal that is not fully
closed, the patient may experience shocking, burning, or irritation at the
neurostimulator implant location, or intermittent stimulation, or loss of
stimulation.