Probe LED Specifi cations:
APPENDIX A Specifi cations
Notices:
Wavelength
660±2nm
1.8mW
2.0mW
940±10nm
RED
IR
Radiant Power
Notes:
●
Speci
fi
cations may be changed without prior notice.
●
The circuit diagrams, the list of components, the illustration of diagrams, and the detailed
rules of calibration, are provided exclusively to professional personnel authorized by our
company.
Specifi cations:
Display
Data: SpO
2
, PR, pulse bar
Others: connection status of probe and other alarm information.
Data update time: less than 5 seconds
Alarm
Alarm:
SpO
2
and pulse rate value, probe off, battery exhausted
Alarm Mode:
audible alarm, visual alarm and information
Alarm Limits range:
SpO
2
70%-100%, PR 30-235 bpm
Default Limits:
SpO
2
High 100%, low 90%; PR High 100 bpm; low 60 bpm
SpO
2
Display Range:
0%~100%
Measurement Range:
70%~100%
Resolution:
1%
Accuracy:
80%~100% : ±2 Digits; 70%~79% : ±3 Digits;
<70% unspeci
fi
ed
Pulse Rate
Display Range:
0~254 bpm
Measurement Range:
30~235 bpm
Resolution:
1 bpm
Accuracy:
±2 bpm or 2% (The larger)
Environment Requirements
Operation Temperature:
5~40 centigrade
Operation Humidity:
≤
80%,no condensation
Storage Temperature:
-20~55 centigrade
Storage Humidity:
≤
93%, no condensation
Power Supply:
3 AA alkaline/Ni-MH batteries or adapter
Working Time:
work for 16 hours continuously at most