Quick Operation Guide |

Positive Control

8

PC

FAM/SYBR

20.715

-

positive (+)

Valid PC

FAM 

and

 HEX 

and

 CY5 have value 

<39

 and 

sigmoidal 

curves

VIC/HEX

30.649

-

positive (+)

CY5

10.649

-

positive (+)

Invalid 

PC

The positive control has failed if 

one of the three channels is negative

 and/or the 

curves are not sigmoidal

 and/or the values are 

≥39

.

Attention:

 However, if a 

sample in the run (FAM) is positive

 despite a failed PC, 

the run must still be scored and the sample result is than 

positive

.

H

  Sample Result Interpretation

Negative Results

21

sample

FAM/SYBR

-

-

negative(-)

Negative

(NA)

HEX value 

<39

 and 

sigmoidal curve

, FAM 

negative

→ 

no virus was amplified.

VIC/HEX

26.317

-

po)

CY5

-

-

po)

Positive Results

21

sample

FAM/SYBR

15.113

-

po)

Positive

Ct 15.113

FAM 

and

 HEX 

and

 CY5 show value 

<39

 and 

sigmoidal 

curve

 structure.

VIC/HEX

21.175

-

po)

CY5

32.175

-

po)

21

sample

FAM/SYBR

15.113

-

po)

Positive

Ct 15.113

No HEX value 

but

 FAM 

and

 CY5 are 

positive

 and show 

sigmoidal curve

, → virus was amplified!

VIC/HEX

-

-

negative(-)

CY5

17.183

-

po)

Invalid Results

21

sample

FAM/SYBR

-

-

negative(-)

Invalid

No Hex value, it is not granted there is enough human 

sample in the tube.

VIC/HEX

-

-

negative(-)

CY5

-

-

negative(-)

Sigmoidal curve

Sigmoidal curve

Non

Summary of Contents for ULTRA SBMS 24

Page 1: ...ULTRA SBMS 24 Quick Operation Guide Real Time qPCR instrument Kurzanleitung Real Time qPCR Ger t Real time qPCR instrument ultraSBMS24 MEDsan Biotech GmbH Brandshofer Deich 10 20539 Hamburg Germany...

Page 2: ...k Operation Guide Manual Packing List Certificate of Conformity B Remove the packing materials and install The Wi Fi antenna Twist the Wi Fi antenna clockwise into the corresponding port as shown in F...

Page 3: ...eriment Figure 4 After setting Load Template please choose the Template MEDsan ul traSBMS SARS CoV 2 and klick Next as shown in the Figure 5 Figure 5 In case you want give the experiment another name...

Page 4: ...HEX and CY5 channel as shown in Figure 8 After the setup is complete klick Ensure Figure 8 1 Barcode scanner mouse and keyboard are optional accessories and not included in the standard product If ev...

Page 5: ...You also can activate both channels at the same time You easily can switch between the overview and the run curve view only Click to Determine and the window change into the sample summary results Her...

Page 6: ...e sample result is than positive H Sample Result Interpretation Negative Results 21 sample FAM SYBR negative Negative NA HEX value 39 and sigmoidal curve FAM negative no virus was amplified VIC HEX 26...

Page 7: ...kliste Konformit ts erkl rung B Entfernen Sie das Verpackungsmaterial und installieren Sie Die Wi Fi Antenne Drehen Sie die Wi Fi Antenne im Uhrzeigersinn in den entsprechenden Anschluss wie in Bild 1...

Page 8: ...Dsan ultraSBMS SARS CoV 2 und klicken Sie auf WEITER wie in der Abbildung 5 gezeigt Bild 5 Wenn Sie dem Experiment einen ande ren Namen geben wollen einen ande ren Namen geben wollen k nnen Sie den Na...

Page 9: ...n Bild 8 Nachdem die Einrichtung abgeschlos sen ist klicken Sie auf BEST TIGEN Bild 8 1 Barcode Scanner Mouse und Tastatur sind optionales Zubeh r und nicht im Standardprodukt enthalten Wenn alles ric...

Page 10: ...h beide Kan le gleichzeitig aktivieren aktivieren Sie k nnen einfach zwischen der bersicht und der reinen Laufkurven ansicht wechseln Klicken Sie auf BESTIMMEN SIE and das Fenster wechselt zu den Erge...

Page 11: ...ser Probe n ist dann positiv H Interpretation der Probenergebnisse Negative Ergebnisse 21 sample FAM SYBR negative Negativ NA HEX Wert 39 und sigmoidal Kurve FAM negativ kein Vieus wurde amplifziert V...

Page 12: ...www medsanbiotech com Support and Sales MEDsan Biotech GmbH Brandshoferdeich 10 20539 Hamburg Deutschland Tel 040 380 238 38 Fax 040 228 585 699 Email info medsanbiotech com MBS24Q0822 v2...

Reviews: