![background image](http://html1.mh-extra.com/html/medsan/ultra-sbms-24/ultra-sbms-24_quick-operation-manual_1762802005.webp)
© MEDsan
®
Biotech GmbH | Brandshofer Deich 10 | 20539 Hamburg | Germany
5
| ULTRA SBMS
24
Professional use only
V2.2 EN 2021-12-16
V2.2 EN 2021-12-16
Quick Operation Guide |
F
General Result Overview
After run is completed, you find the general
overview or your results.
You can activate/deactivate
FAM , HEX
or
CY5
channel
In the general overview you easy can proof
if the relevant curves are Sigmoidal or not.
You can activate/deactivate
the related
channel.
You also can
[
activate
]
both channels at
the same time.
You easily can switch between
the
overview
and the
run
curve view
only
Click to
[
Determine
]
and the window
change into the sample summary results.
Here you will find
all channels for each
sample
as well your Controls such as NC
and PC.
21
sample
FAM/SYBR
-
-
negative(-)
VIC/HEX
-
-
negative(-)
CY5
-
-
negative(-)
G
Validation of Negative and Positive Control
Negative Control
17
NC
FAM/SYBR
-
-
negative(-)
Valid NC
No values are shown up. Run is valid if PC also is valid.
VIC/HEX
-
-
negative(-)
CY5
-
-
negative(-)
Invalid
NC
If
one
of the two channels
has values
, the NC
has failed
, the run is considered
„
invalid
“ and must be repeated.