Routine Operation
54
03-11-24
1005en01
The background should not be higher than the figures below, assuming that at
least 2 blanks are run after a sample.
( CGS units) (“Units =1”:
7.2
Analysing the Sample (Open Tube)
Choose the operational mode with [Menu-Operate] so that the last run sample is
displayed.
Aspirate the sample through the aspirating pipette by pressing the start lever be-
hind the aspiration needle. See picture below.
The display shows the following sequences:
RBC
0.01
WBC
0.1
HGB
0.1
PLT
10
1052en.gif
1053.bmp
Warning
Contamination hazard if
contaminated blood enters
into open cut.
Always wear protective
gloves when handling
blood samples.
(Gloves in the picture are
not shown for clarity rea-
sons only.)
Last Sample (= blank)
Aspirating sequence
1055en.gif
1052en.gif