Installation
46
03-11-24
1003en01
Note:
Aspiration command keys are not located on the keyboard. Aspiration of a sam-
ple is started by pressing the start-lever behind the aspiration needle.
5.5
Filling the system with reagents
After power-on, the instrument will perform a self-check which lasts for ca. 2
minutes. During this period, no key entry is accepted by the instrument.
1.
Press
[
Menu
]
until the main menu is displayed on the LCD screen.
2.
Scroll with the up-arrow key to 8 and press
[
Enter
].
3.
Scroll to 8.2 and press
[
Enter
]
.
The system is now filled with reagents. This cycle lasts for ca. 6 minutes.
When the system is finished with this cycle, the cursor field returns to the field
8.2. Press
[
Menu
]
to return to the main menu.
1037en.gif
1036en.gif