• Using 

the 

Rewarm Patient From 

temperature, the 

Rewarm Patient To 

temperature and the rewarming rate settings, the system will calculate 
and display the rewarming duration and the date/time at which the the 
patient will reach the final rewarming target temperature.

• Press 

the 

Save

 button to save the new settings and close the

 Rewarm 

Patient-Adjust 

window. 

3. Initiate Patient Cooling 

• Press 

Start

, in the 

Cool Patient

 window to initiate therapy. You will hear 

a tone and then a voice stating “Therapy Started”. Additionally, the 

Cool 

Patient

 window and the 

A

RCTIC

 S

UN

®

 Temperature Management System 

icon will blink, indicating that therapy is in progress. 

4. Initiate Patient Rewarming 

•  Upon completion of the cooling phase, there are two options for initia-

tion of patient rewarming, either Automatically or Manually, depending 
on the 

Rewarming Begins 

setting in 

Hypothermia Settings

.  

• If 

Rewarming Begins 

is set to 

Automatically

, the rewarming process 

starts automatically when the 

Cool Patient 

therapy is complete and the 

duration reaches zero.

• If 

Rewarming Begins

 is set to 

Manually

, the rewarming process starts 

when the 

Start

 button is pressed in the 

Rewarm Patient 

window.  The 

cooling process will continue until the 

Rewarm Patient Start 

button is 

pressed.  An 

Alert

 will occur when the 

Cool Patient 

duration reaches 

zero. 

End Therapy 

• From 

the 

Normothermia

 therapy or 

Hypothermia

 therapy screen, press 

the 

Stop

 button to terminate water circulation to the pads. 

• Press 

the 

Empty Pads

 button, and follow the instructions to purge the 

pads of water. 

•  Disconnect the pads from the Fluid Delivery Line. 
•  Slowly and carefully remove pads from the patient skin.
•  Discard the used pads in accordance with hospital procedures for 

medical waste. 

•  Press the power switch 

Off

If power is lost while the power switch is in the On position, an audible alert 
will be issued until it is switched Off.  This alerts the user that the treatment 
may have been accidentally stopped.

Chapter 3 – Normothermia Settings 

Normothermia Settings 

Use the 

Normothermia Settings 

screen to view the current settings and 

modify the settings for the following parameters.  To modify any parameter 
setting, press the 

Adjust

 button to the right of the parameter. 

Normothermia Settings screen parameters: 

Therapy Settings 

• Timer 

Begins 

Water Temperature Settings 

• Pre-Condition 

Water 

• Manual 

Control 

•  High Water Limit 
•  Low Water Limit 

Patient Temperature Settings 

•  High Patient Alert 
•  Low Patient Alert 
• Control 

Strategy 

Display Settings 

• Temperature 

Units 

•  Temperature Units Adjust 
•  Patient Temp 2 
• Speaker 

Volume

To access the Normothermia Settings screen: 

1) Press 

Adjust

 on the 

Control Patient 

window. 

2) Press 

the 

More

 button on the 

Control Patient Adjust 

window. 

3) The 

Normothermia Settings

 screen will be displayed.  

4)  To save the new settings as the current patient therapy settings, press 

the 

Close

 button. For instructions on saving the settings as the system 

defaults, see 

Advanced Setup

O P E R A T O R ’ S   M A N U A L

Summary of Contents for ARCTIC SUN 5000

Page 1: ...O P E R AT O R S M A N U A L ...

Page 2: ...Hypothermia Cool Patient and Rewarm Patient 5 End Therapy 6 Chapter 3 Normothermia Settings 6 Normothermia Settings 6 Chapter 4 Hypothermia Settings 7 Hypothermia Settings 7 Chapter 5 Advanced Setup 7 Chapter 6 Alarms and Alerts 8 Alarms 8 Main Safety Alarms 8 Non Recoverable Alarms 8 Recoverable Alarms 8 Alerts 8 Chapter 7 Maintenance and Service 8 Cleaning and Maintenance 8 Software Update 9 Ser...

Page 3: ...ne check connections If needed replace the leaking pad Leakage may result in lower flow rates and potentially decrease the performance of the system The ARCTIC SUN Temperature Management System is for use only with the ARCTICGEL Pads The ARCTICGEL Pads are only for use with the ARCTIC SUN Temperature Management Systems The ARCTICGEL Pads are non sterile for single patient use Do not reprocess or st...

Page 4: ...Fluid Delivery Line Patient Temp 1 cable Patient Temp 2 cable optional and Fill Tube to the back of the control module 2 Plug the Power Cord into the wall outlet Position ARCTIC SUN Temperature Man agement System so that access to the power cord is not restricted Power On 1 Turn the power ON by activating the Power Switch 2 The control module will automatically go through a brief self test of the ...

Page 5: ...ry Line While holding the pad line tubing insert the clear pad line connector into the Fluid Delivery Line manifold Do not press or squeeze the wings when connecting The connector will click into place Temperature Probe Placement Patient temperature control with the ARCTIC SUN Temperature Management System requires patient temperature feedback provided by an indwelling patient temperature probe co...

Page 6: ...ol Patient window in the Normothermia therapy screen The Control Patient window displays the patient target temperature and the duration since the initiation of normothermia therapy To initiate Normothermia therapy 1 From the Patient Therapy Selection screen press the Normothermia button to display the Normothermia therapy screen 2 The default patient target temperature will display in the Control...

Page 7: ...cooling duration to cool the patient before rewarming begins Press the Save button to save the new settings and close the Cool Patient Adjust window 2 Rewarm Patient Settings The default patient target temperature and duration will display in the Rewarm Patient window To change the rewarming phase patient target temperature and rewarming rate press the Adjust button in the Rewarm Patient window to...

Page 8: ...apy screen press the Stop button to terminate water circulation to the pads Press the Empty Pads button and follow the instructions to purge the pads of water Disconnect the pads from the Fluid Delivery Line Slowly and carefully remove pads from the patient skin Discard the used pads in accordance with hospital procedures for medical waste Press the power switch Off If power is lost while the powe...

Page 9: ...rapy settings press the Close button For instructions on saving the settings as the system defaults see Advanced Setup Chapter 5 Advanced Setup Use the Advanced Setup screen to view the current settings and modify the settings for the following parameters To modify any parameter setting press the Adjust button to the right of the parameter Location Time Settings Language Number Format Current Time...

Page 10: ...ype of Alarm is known as Non Recoverable If this situation occurs cycle the device power turn device Off then On If the alarm recurs contact Medivance Customer Support Recoverable Alarms Other Alarms that temporarily Stop the device until the user is able to correct the cause and clear the Alarm are classified as Recoverable If the condition that initiated the alarm is not addressed and problem per...

Page 11: ...Customer Support for technical support and customer service instructions to enable appropriately qualified technical personnel to repair those parts of the equipment that Medivance considers repairable Calibration See ARCTIC SUN 5000 Temperature Management System Service Manual for calibration requirements and instructions Appendix A Product Specifications Technical Description The ARCTIC SUN Temper...

Page 12: ...d from testing in simulated use Patient Temperature Control Range 32 C to 38 5 C 89 6 F to 101 3 F 0 1 C F increments Water Temperature Display Range 3 C to 45 C 37 4 F to 113 0 F 0 1 C F increments Water Temperature Control Range Manual 4 C to 42 C 39 2 F to 107 6 F 1 C F increments High Water Temperature Limit 36 C to 42 C 96 8 F to 107 6 F 1 C F increments Low Water Temperature Limit 4 C to 25 ...

Page 13: ...with CAN CSA C22 2 STD 601 1 M90 and UL STD 6060 1 Indicates high temperature part or component Indicates that only sterile or distilled water should be used when filling the ARCTIC SUN Temperature Management System Control Module Identifies Patient Temperature 1 the patient temperature probe input for monitoring and control Identifies Patient Temperature 2 the patient temperature probe input for mon...

Page 14: ...er emissions IEC 61000 3 3 Complies 1 2 EN IEC 60601 1 2 Table 2 Guidance and Manufacturer s Declaration Electromagnetic Immunity The ARCTIC SUN 5000 Temperature Management System unit is intended for use in the electromagnetic environment specified below The customer or the end user of the ARCTIC SUN 5000 Temperature Management System unit should assure it is used only in such an environment Immun...

Page 15: ... If the measured field strength in the location in which the ARCTIC SUN 5000 Temperature Management System unit is used exceeds the applicable RF compliance level above the ARCTIC SUN 5000 Temperature Management System unit should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the ARCTIC SUN 5000 Tem...