2

• 

The USB data port is to be used only with a standalone USB flash drive. Do not 
connect to another mains powered device during patient treatment. 

• 

Users should not use cleaning or decontamination methods different from those 
recommended by the manufacturer without first checking with the manufacturer that 
the proposed methods will not damage the equipment. Do not use bleach (sodium 
hypochlorite) as it may damage the system. 

• 

Medivance will not be responsible for patient safety or equipment performance if the 
procedures to operate, maintain, modify or service the Medivance 

A

RCTIC

 S

UN

®

 Tem-

perature Management System are other than those specified by Medivance. Anyone 
performing the procedures must be appropriately trained and qualified. 

System Setup 

Unpack 

1) Unpack 

the 

A

RCTIC

 S

UN

®

 Temperature Management System Control Module and 

accessories. 

2)  Allow the control module to remain upright for at least 2 hours prior to completing the 

installation and setup procedure in order to allow the chiller oil to settle.  Damage to 
the chiller compressor may result otherwise. 

Connections 

1)  Use only Medivance cables and accessories with the 

A

RCTIC

 S

UN

®

 Temperature 

Management System Control Module. Connect the Fluid Delivery Line, Patient 
Temp 1 cable, Patient Temp 2 cable (optional) and Fill Tube to the back of the 
control module. 

2)  Plug the Power Cord into the wall outlet. Position 

A

RCTIC

 S

UN

®

 Temperature Man-

agement System so that access to the power cord is not restricted.

Power On 

1)  Turn the power ON by activating the Power Switch. 
2)  The control module will automatically go through a brief self-test of the 

independent safety alarm. 

3)  A New User Training module option is available from the start up 

screen. 

 

4)  When the self-test is complete, the 

Patient Therapy Selection

 screen 

will appear on the control panel. 

Fill Reservoir 

1)  Fill the reservoir with sterile or distilled water only. 
2)  Four liters of water will be required to fill the reservoir at initial installa-

tion. 

3)  Add one vial of 

A

RCTIC

 S

UN

®

 Temperature Management System Cleaning 

Solution to the sterile or distilled water. 

4) From 

the 

Patient Therapy Selection

 screen, press either the 

Normo-

thermia

 or 

Hypothermia

 button, under the New Patient heading. 

5) From 

the 

Hypothermia

 or 

Normothermia

 therapy screen, press the 

Fill 

Reservoir

 button. 

6) The 

Fill Reservoir 

screen will appear.  Follow the directions on the 

screen. 

 

Power Switch

Patient
Temp
Out

Fluid 
Delivery 
Line

Power Cord 

Equipotential
Stud

Patient
Temp 1

Strain Relief

Patient
Temp 2

Patient

Temp Cable

Fill Tube 

Storage

Fill Tube/Port

Drain Ports

Air Filter

O P E R A T O R ’ S   M A N U A L

Summary of Contents for ARCTIC SUN 5000

Page 1: ...O P E R AT O R S M A N U A L ...

Page 2: ...Hypothermia Cool Patient and Rewarm Patient 5 End Therapy 6 Chapter 3 Normothermia Settings 6 Normothermia Settings 6 Chapter 4 Hypothermia Settings 7 Hypothermia Settings 7 Chapter 5 Advanced Setup 7 Chapter 6 Alarms and Alerts 8 Alarms 8 Main Safety Alarms 8 Non Recoverable Alarms 8 Recoverable Alarms 8 Alerts 8 Chapter 7 Maintenance and Service 8 Cleaning and Maintenance 8 Software Update 9 Ser...

Page 3: ...ne check connections If needed replace the leaking pad Leakage may result in lower flow rates and potentially decrease the performance of the system The ARCTIC SUN Temperature Management System is for use only with the ARCTICGEL Pads The ARCTICGEL Pads are only for use with the ARCTIC SUN Temperature Management Systems The ARCTICGEL Pads are non sterile for single patient use Do not reprocess or st...

Page 4: ...Fluid Delivery Line Patient Temp 1 cable Patient Temp 2 cable optional and Fill Tube to the back of the control module 2 Plug the Power Cord into the wall outlet Position ARCTIC SUN Temperature Man agement System so that access to the power cord is not restricted Power On 1 Turn the power ON by activating the Power Switch 2 The control module will automatically go through a brief self test of the ...

Page 5: ...ry Line While holding the pad line tubing insert the clear pad line connector into the Fluid Delivery Line manifold Do not press or squeeze the wings when connecting The connector will click into place Temperature Probe Placement Patient temperature control with the ARCTIC SUN Temperature Management System requires patient temperature feedback provided by an indwelling patient temperature probe co...

Page 6: ...ol Patient window in the Normothermia therapy screen The Control Patient window displays the patient target temperature and the duration since the initiation of normothermia therapy To initiate Normothermia therapy 1 From the Patient Therapy Selection screen press the Normothermia button to display the Normothermia therapy screen 2 The default patient target temperature will display in the Control...

Page 7: ...cooling duration to cool the patient before rewarming begins Press the Save button to save the new settings and close the Cool Patient Adjust window 2 Rewarm Patient Settings The default patient target temperature and duration will display in the Rewarm Patient window To change the rewarming phase patient target temperature and rewarming rate press the Adjust button in the Rewarm Patient window to...

Page 8: ...apy screen press the Stop button to terminate water circulation to the pads Press the Empty Pads button and follow the instructions to purge the pads of water Disconnect the pads from the Fluid Delivery Line Slowly and carefully remove pads from the patient skin Discard the used pads in accordance with hospital procedures for medical waste Press the power switch Off If power is lost while the powe...

Page 9: ...rapy settings press the Close button For instructions on saving the settings as the system defaults see Advanced Setup Chapter 5 Advanced Setup Use the Advanced Setup screen to view the current settings and modify the settings for the following parameters To modify any parameter setting press the Adjust button to the right of the parameter Location Time Settings Language Number Format Current Time...

Page 10: ...ype of Alarm is known as Non Recoverable If this situation occurs cycle the device power turn device Off then On If the alarm recurs contact Medivance Customer Support Recoverable Alarms Other Alarms that temporarily Stop the device until the user is able to correct the cause and clear the Alarm are classified as Recoverable If the condition that initiated the alarm is not addressed and problem per...

Page 11: ...Customer Support for technical support and customer service instructions to enable appropriately qualified technical personnel to repair those parts of the equipment that Medivance considers repairable Calibration See ARCTIC SUN 5000 Temperature Management System Service Manual for calibration requirements and instructions Appendix A Product Specifications Technical Description The ARCTIC SUN Temper...

Page 12: ...d from testing in simulated use Patient Temperature Control Range 32 C to 38 5 C 89 6 F to 101 3 F 0 1 C F increments Water Temperature Display Range 3 C to 45 C 37 4 F to 113 0 F 0 1 C F increments Water Temperature Control Range Manual 4 C to 42 C 39 2 F to 107 6 F 1 C F increments High Water Temperature Limit 36 C to 42 C 96 8 F to 107 6 F 1 C F increments Low Water Temperature Limit 4 C to 25 ...

Page 13: ...with CAN CSA C22 2 STD 601 1 M90 and UL STD 6060 1 Indicates high temperature part or component Indicates that only sterile or distilled water should be used when filling the ARCTIC SUN Temperature Management System Control Module Identifies Patient Temperature 1 the patient temperature probe input for monitoring and control Identifies Patient Temperature 2 the patient temperature probe input for mon...

Page 14: ...er emissions IEC 61000 3 3 Complies 1 2 EN IEC 60601 1 2 Table 2 Guidance and Manufacturer s Declaration Electromagnetic Immunity The ARCTIC SUN 5000 Temperature Management System unit is intended for use in the electromagnetic environment specified below The customer or the end user of the ARCTIC SUN 5000 Temperature Management System unit should assure it is used only in such an environment Immun...

Page 15: ... If the measured field strength in the location in which the ARCTIC SUN 5000 Temperature Management System unit is used exceeds the applicable RF compliance level above the ARCTIC SUN 5000 Temperature Management System unit should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the ARCTIC SUN 5000 Tem...