The USB data port is to be used only with a standalone USB flash drive. Do not
connect to another mains powered device during patient treatment.
•
Users should not use cleaning or decontamination methods different from those
recommended by the manufacturer without first checking with the manufacturer that
the proposed methods will not damage the equipment. Do not use bleach (sodium
hypochlorite) as it may damage the system.
•
Medivance will not be responsible for patient safety or equipment performance if the
procedures to operate, maintain, modify or service the Medivance
A
RCTIC
S
UN
®
Tem-
perature Management System are other than those specified by Medivance. Anyone
performing the procedures must be appropriately trained and qualified.
System Setup
Unpack
1) Unpack
the
A
RCTIC
S
UN
®
Temperature Management System Control Module and
accessories.
2) Allow the control module to remain upright for at least 2 hours prior to completing the
installation and setup procedure in order to allow the chiller oil to settle. Damage to
the chiller compressor may result otherwise.
Connections
1) Use only Medivance cables and accessories with the
A
RCTIC
S
UN
®
Temperature
Management System Control Module. Connect the Fluid Delivery Line, Patient
Temp 1 cable, Patient Temp 2 cable (optional) and Fill Tube to the back of the
control module.
2) Plug the Power Cord into the wall outlet. Position
A
RCTIC
S
UN
®
Temperature Man-
agement System so that access to the power cord is not restricted.
Power On
1) Turn the power ON by activating the Power Switch.
2) The control module will automatically go through a brief self-test of the
independent safety alarm.
3) A New User Training module option is available from the start up
screen.
4) When the self-test is complete, the
Patient Therapy Selection
screen
will appear on the control panel.
Fill Reservoir
1) Fill the reservoir with sterile or distilled water only.
2) Four liters of water will be required to fill the reservoir at initial installa-
tion.
3) Add one vial of
A
RCTIC
S
UN
®
Temperature Management System Cleaning
Solution to the sterile or distilled water.
4) From
the
Patient Therapy Selection
screen, press either the
Normo-
thermia
or
Hypothermia
button, under the New Patient heading.
5) From
the
Hypothermia
or
Normothermia
therapy screen, press the
Fill
Reservoir
button.
6) The
Fill Reservoir
screen will appear. Follow the directions on the
screen.
Power Switch
Patient
Temp
Out
Fluid
Delivery
Line
Power Cord
Equipotential
Stud
Patient
Temp 1
Strain Relief
Patient
Temp 2
Patient
Temp Cable
Fill Tube
Storage
Fill Tube/Port
Drain Ports
Air Filter