UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
9 / 15
0 / I
Switch On-Switch Off
Control Panel
Turn the right - the vacuum is rising
Control Panel
Attention, see the User’s Manual
Control Panel
Foot Control
Electric Label
Fuse
Electric Label
5. Unit interconnection
5.1. Interconnection of the unit with basic equipment
Procedure:
1. Connect the silicone hose (2) to the hose
connector VACUUM (1).
2. Connect the bacteriologic filter (3) to the hose
(2). The side of filter marked „IN“ was directed
to the safety jar (6).
3. Shift the hose (4) onto the other bacteriologic
filter connector.
4. Shift the hose (4) to the output connector (5)
(safety valve under the connector is safety
valve (7)) of safety jar.
5. Shift the hose (10) on the input connector (9)
(stream-line tip (8) is under connector) of safety
jar (6).
6. Shift the loose end of the hose (10) on the
output connector (12) of collecting jar - a safety
valve is under connector - it is labelled as
“VACUUM”.
7. Shift the suction hose (11) on the input
connector (13) (without safety valve) of the
collecting jar - it is labelled as “PATIENT”.
6. Switching on and operation
Attention:
Ensure so as before each suction by new patient was realized sterilization of suction
hose with suction adapter, vessel for secretion, safety vessel and interconnection hose to
microbiological filter.
6.1. Suction unit operational set-out
Connect a suction unit with a vessel exit connector to secret – it is labelled as “VACUUM”. Hoses must not be
damaged, cracked or messed up.
Important!!!
Check the bacteriological filter from side “IN” and if the filter is glazed replace its.
6.2. Vessel preparation for secret
Vessels must be controlled for possible breaks, granularity or damage before use. Any defect can produce leakage or
split of the vessel (implosion of the vessel on secret is possible at vacuum).
Before sucking we recommend to infuse into the secret vessel a little water or disinfecting agent for easier cleaning.