Medist MEVACS M90 User Manual Download Page 10

 

 

 

 

 

 

 
UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009 

 

 

T-QN-04/05-01-1

 

 

10 / 15 

6.3. Switching on 

1.  The unit is switched on by means of ON-OFF switch. 

  position „0“ - the unit is off 

  position „I“ - the unit is on (the switch illuminates) 

 
2.  Suction circuit  - option and switching on  
 
You can choose left or right suction circuit. It is possible to use both suction circuits at the same time. 
 
 
 
 

 

 

 

 
 
3.  Select the respective required value by regulator valve. 
 
Vacuum is increasing by turning round of a regulating vent with the sun. Value of vacuum is displayed only if filtering 
perimeter is closed, e.g. by cranking the end of the filtering hose.

  

 

 

 

 
Vacuum-meter scale is into bars. 
 
-1bar  =  -100 kPa  =  -100% vacuum 

 

6.4. Functional examination of suction unit 

6.4.1. Inner suction circuit 

 

turn the suction unit on 

 

cover the connector (1) on the suction unit by your finger 

 

turn the vacuum regulating vent with the sun till it stop 

 

maximum vacuum value must be displayed on the vacuum-meter (see point “Technical data”) 

 
Status of components and leakage inside of suction unit is controlled by the test. 

6.4.2. Outer suction circuit 

 

turn the suction unit on 

 

crank or cover free end of the suction hose (11) by your finger 

 

turn the vacuum regulating vent with the sun till it stop 

 

vacuum-meter  must  display  maximum  value  of  vacuum  (see  point  “Technical  data”)  after 
some time (it depends of the secret vessel volume) 

 
Status of components and leakage inside of suction unit outside its own body is controlled by the test. 

Suction circuit off 

Suction circuit on 

VACUOMETER SCALE 

Summary of Contents for MEVACS M90

Page 1: ...IST s r o Petrusovskeho 4 066 01 Humenne Slovakia T Fax 00421 57 7753276 E mail medist medist sk http www medist sk T QN 04 05 01 1 2070 Medical electric suction unit Ordering number MEVACS M90 72001...

Page 2: ...service extension are made by MEDIST company or authorized service organizations the device is used in accordance with this instruction manual Use only microbiological filters supplied by manufacture...

Page 3: ...lectric mains when unplugged from the unit socket Attendant must make certain before every use of suction unit about its good condition and device ready Don t expose the collection vessels for direct...

Page 4: ...H BASIC EQUIPMENT 9 6 SWITCHING ON AND OPERATION 9 6 1 SUCTION UNIT OPERATIONAL SET OUT 9 6 2 VESSEL PREPARATION FOR SECRET 9 6 3 SWITCHING ON 10 6 4 FUNCTIONAL EXAMINATION OF SUCTION UNIT 10 6 4 1 In...

Page 5: ...drophobic bacteriological filter 11 Protection of the unit personnel and patient against infection by means of fitted bacteriologic filter 12 The possibility of all functions operation by means of foo...

Page 6: ...ssure fit lid and metal tubing nipple 42011 12 15 4 0 L jar TM polysulfone with pressure fit lid and metal tubing nipple 42011 10 16 MONOVAC suction liner capacity 1 4 L 42002 01 17 EURO universal rai...

Page 7: ...x 2 0Lit jar MEVACS M90 2 x 4 0Lit jars 2 x 2 0Lit jars MEVACS M90 2 x 4 0Lit jars 1 x 2 0Lit jar foot switch 3 Installation The MEVACS suction unit is a compact unit There is no needed assembly 3 1 U...

Page 8: ...els with brake 8 foot control 9 change valve 10 microbiologic hydrophobic filter 11 EURO rail adapter 12 safety jar 13 safety overflow valve 14 hose connector for connection of vacuum branch 15 safety...

Page 9: ...safety valve is under connector it is labelled as VACUUM 7 Shift the suction hose 11 on the input connector 13 without safety valve of the collecting jar it is labelled as PATIENT 6 Switching on and...

Page 10: ...acuum 6 4 Functional examination of suction unit 6 4 1 Inner suction circuit turn the suction unit on cover the connector 1 on the suction unit by your finger turn the vacuum regulating vent with the...

Page 11: ...removed during clearing 6 6 Changing of bacteriologic filter Suction microbiologic filter eliminated risk of contamination with bacteria viruses and infected particle from patient to suction pump or c...

Page 12: ...2 x 2 0 4 0 LT polycarbonate jars with overflow valve Suction Suction tube Silicone 8 x 3 mm 2 0 m Operation Continuous Filtration Hydrophobic suction microbiological filter with very high bacterial...

Page 13: ...protect together suction unit for overflow secretion The plastic case holders truck and other equipment can be cleaned with soft wet cloth moistened by cleaning agent which does not react with the pl...

Page 14: ...ied faults contact either qualified persons of service company or directly the manufacturer MEDIST s r o Petrusovskeho 4 066 01 Humenne Slovakia tel fax 00421 57 7753276 E mail medist medist sk 8 3 Re...

Page 15: ...2 0 L jar TM with screw metal lid polycarbonate autoclavable 121 O C No 42011 14 polysulfone autoclavable 134 O C No 42011 21 safety valve in lid metal connectors 4 0 L jar TM with pressure fit lid p...

Reviews: