21
parameters are adhered to during every cycle. Also observe the
applicable legal requirements in your country as well as the hygiene
regulations of the hospital or clinic, especially with regard to the
additional requirements for the inactivation of prions.
4.1.2
Post-operation treatment
The post-operation treatment must be carried out immediately, no
later than 30 minutes after the completion of the operation. The steps
are as follows:
a.
Let the device works for 30 seconds under cleaning mode (pg. 19)
to flush the system internally, the handpiece and the attached tip;
b.
Remove the handpiece from the main unit, and rinse away the dirt
on the surface of product with water (distilled water/ deionized water);
c.
Dry the product with a clean, soft cloth and place it in a clean tray.
WARNING:
The water to be used to clean must be pure water, distilled water or
deionized water.
4.2
Preparation before cleaning
Tools: Torque wrench, tray, soft brush, clean and dry soft cloth.
a.
Remove the tip from handpiece with torque wrench (or wrench)
provided, and then put the tip and wrench into a clean tray.
b.
Unscrew the nipple of product counterclockwise, remove the
sealing ring, light pipe, and LED lamp, and put them in the tray.
c.
Use a clean soft brush to carefully brush the joints between product
and the connector of cable, front thread, horn, nipple, seal ring, light
pipe and LED lamp until the dirt on surface is not visible. Then use soft
cloth to dry the product and accessories and put them into a clean
tray. The cleaning agent can be pure water, distilled water or
deionized water.
Disassembling parts