19
3.2
Corrective maintenance
In the event of faulty operation, the following corrective maintenance
actions may be performed by the user.
3.2.1
Clean the water pipe with the air-water syringe (Fig. 11):
Fig. 11. Water line at the back of the unit.
a. Snip the water pipe at a distance of 5 to 7 inches from the
external water connection.
b. Turn on the power switch.
c. Adjust the tip of the air-water syringe to the water pipe.
d. Screw off the tip or pull out the handpiece.
e. Step on the foot switch.
f. Turn on the air-water syringe pressing air or water into the water
pipe to unblock the system.
3.3
Cleaning mode
It is important to purge the system after use preventing corrosion or
mineral deposits alongside the internal system.
3.3.1
Operation:
a. After finishing operation, remove the reservoir and dispose the
remaining irrigant. Rinse the tank with water. Fill the tank with distilled
water to a quarter of its capacity. Reinstall the tank. Press
simultaneously
the
two buttons on the touch
screen panel (water
line mode and reservoir
mode buttons, Fig. 6).