![MEDICSIGHT MEDICCO2LON User Manual Download Page 31](http://html.mh-extra.com/html/medicsight/medicco2lon/medicco2lon_user-manual_1761096031.webp)
31
Copyright © 2020 Ultrasound Technologies Ltd
31
7.1
MEDICCO
2
LON Device
Control
EU/USA Version
Japanese Version
Target Pressure
5 - 30 mmHg selectable in 1mmHg steps
Pressure Hold
±10% of Target Pressure
Target Flow 1, 2, 3, 4 LPM selectable
1, 2, 3 LPM
Flow Profile
0.5LPM per 0.5L delivered volume, up to target flow rate
Volume
12L maximum delivery with optional volume stops at 4, 6, 8 and 10L
Display
Range
Accuracy
Gas Pressure 0 - 75 mmHg
±1mmHg of displayed value
Gas Flow 0 - 5 LPM
±10% at 3LPM
Gas Volume 0 - 12 L
±20% of displayed value
Pressure Relief
Activation Pressure
Accuracy
Electronic Valve 50mmHg or
+10% of Target Pressure for 5 seconds
±1 mmHg
Mechanical Valve 1.5psi (78mmHg nominal)
±20%
Physical
Size 235 mm x 160 mm x 180 mm (width x height x depth)
Weight <2.0 Kg
Data Presentation 150mm x 90mm LED Backlit LCD with Resistive Touch Interface
CO
2
Supply
CO
2
Supply Pressure 1.5 - 2.0 Bar Regulated
Gas Supply Single “C” or “D” size CO
2
cylinder
Electrical Power Supply
AC Input Voltage 90 - 260 VAC
Frequency 46-64Hz
Power Consumption 40VA
Fuse Type/Rating High Breaking Capacity T3.15A 250V
Safety
Standards Designed and tested to BS EN 60601-1:2006
Electrical Designation Class I Type B
7.2
Administration Set
Single use set including catheter, syringe, tubing, connectors and 0.1 micron
PTFE porous hydrophobic filter.
7. EQUIPMENT
SPECIFICATION