4. INSTALLATION
and SETUP
4.1.3 Insufflator Device Preparation
Electrical Connections
The following procedures should be performed before each use of the device:
• Visually inspect the electrical connections. Do not use if inspection reveals any
damage.
• Connect the power cable to the power inlet connection on the back of the
MedicCO
2
LON device.
• Ensure the main power switch is off and that the voltage is correct when
connecting the power cord to the wall outlet.
4.1.4 CO
2
Gas Connections
Connecting the Regulator to the Cylinder
Medical Grade gases are supplied in cylinders that are fitted with delivery valves. There
are two popular versions of valve:
PIN Index – often used as the preferred valve for Medical Grade gases.
Bull Nose – common in industrial and laboratory applications.
MedicCO
2
LON is supplied with a Pressure Regulator that can be connected to a Bull Nose
valve directly.
To connect Bull Nose valve cylinders:
1. Close the regulator by rotating the valve handle counter-clockwise.
2. Connect the regulator to the cylinder using the spanner provided and securely
tighten.
3. Attach the connection tube supplied to the outlet connection and attach the
other end of the tube to the CO
2
input connection on the MedicCO
2
LON device.
4. Slowly open the gas cylinder valve. A leaking valve can be identified by listening
for any seepage sound. A leaking valve should be reported immediately as it can
be a serious safety issue.