4.3.1 Workflow Setup
The ‘Workflow’ setup screen allows the user to modify the default target pressure icons
and activate or deactivate the automatic volume stop function.
4.3.1.1
Modify Pressure Target icons
1. Select the setup icon on the main screen.
2. Select a preset target pressure (1), this will highlight blue.
3. Modify using the increment and decrement icons (3).
4. Press the accept icon to confirm and exit the setup screen.
5. Press the decline icon to exit without saving changes.
4.3.1.2 Automatic Volume Stop
The Volume Stop function pauses the insufflation of gas at the following volumes; 4,6,8
and 10L, once a volume stop level is reached the unit will stop insufflating and an alarm
will be sounded. Press the resume icon to continue insufflating.
1. Select the setup icon on the main screen.
2. Press the volume stop icon to activate or deactivate (4).
Active
Inactive
3. Press the accept icon to confirm and exit the setup screen.
4. Press the decline icon to exit without saving changes.
N.B - The current set up target values and volume stop sequence will be stored when the
MedicCO
2
LON device is shutdown.
4.3.1.3 Return to Factory Setting
1. Press and hold the factory settings icon (6) until all pressure preset values turn
blue. This indicates that they have been restored to factory default values.
2. Press the accept icon to confirm and exit the setup screen.
3. Press the decline icon to exit without saving changes.