and the total gas volume shown on the display flashes. These indications signify
that the first phase of CO
2
insufflation is over.
N.B – Volume stop must be activated for the above to occur. (see section 4.3.1.2)
12. The second phase of CO
2
insufflation can be started by pressing ‘resume’. CO
2
insufflation resumes until the total CO
2
insufflated volume reaches 6.0 litres.
13. The CO
2
insufflation processes can similarly be resumed at 8.0 litres and 10.0
litres. When the total CO
2
insufflated gas volume reaches 12 litres, the process is
terminated. The process cannot be continued by pressing ‘RESUME’. Total Volume
Display changes to red and a buzzer continues to sound, this can be silenced by
selecting the stop icon.
14. To reset the volume display and open the gas relief valves, hold the ‘RESET’ icon
for two seconds. The indication of the total CO
2
volume used reverts to 0.0L. All
other internal parameters related to the process are returned to default settings.
N.B – The reset icon is inactive during insufflation
5.1
Shut-Down Procedure
ATTENTION:
Ensure the CO
2
cylinder valve is closed when switching off the system.
1. Disconnect the gas administration tubing from the front panel and let the gas
discharge to the atmosphere.
2. Using the syringe provided deflate the rectal catheter balloon (if inflated).
3. Remove the rectal catheter carefully from the patient and dispose of the
administration set.
4. Switch off the electrical supply to the machine using the Power ON/OFF Switch
located on the Rear Panel of the machine and disconnect the power cord from
the power outlet.