6. Maintenance Intervals
Inspection and maintenance
Once a year or after max. 2800 working
hours by the service technician responsible
* Coarse dust filter
Clean filter fleece once a week. If blocked,
exchange by the service technician responsible.
* Fine filter
Exchange every six month or after max. 2000 working hours
by the service technician responsible.
* Microfilter
Exchange every six month or after max. 2000 working hours
by the service technician responsible
Patient bacteria filter
Exchange during inspection and maintenance by the
service technician responsible
Inhalation equipment (e.g. nasal canula)
Exchange according to instructions of
and oxygen extension hose
the attending physician
* Note! All maintenance intervals are guidelines. In different cases these can be shorter due to changing
conditions regarding temperature, humidity, dust etc. The filter service life is 20% less in devi-
ces with 6 l capacity.
7. Specifications
Model (PRECISE 6000) M
Model (PRECISE 6000) M S
Ambient temperature running:
10 to 40° C
10 to 40° C
Dimensions:
55 x 21.5 x 55 cm
55 x 21.5 x 55 cm
Weight:
approx. 21 kg
approx. 21 kg
Line voltage:
230 V +/-10%
230 V +/-10%
50 Hz
50 Hz
Nominal current:
2.0 A
2.0 A
Flow:
0.1 - 5 l/min
0.1 - 6 l/min
Oxygen concentration
0.1 - 3 l/min
0.1 - 4 l/min
at 20°C, 1013 hPa
95 +/- 3 Vol.% O2
95 +/- 3 Vol.% O2
50% rel. humidity
-4 l/min
-5 l/min
90 +/- 3 Vol.% O2
88 +/- 3 Vol.% O2
- 5 l/min
- 6 l/min
82 +/- 3 Vol.% O2
80 +/- 3 Vol.% O2
Sound pressure level
38 dB(A)
39 dB(A)
Output pressure
300-350 mbar
300-350 mbar
17