15
2.6. Data Interface Option D
By means of the cable supplied the concentrator
can be conntected to a computer with SB-interfa-
ce. This makes the concentrator’s service easier.
Apart form that the interface serves as a connec-
tion to a low consumption oxygen monitor.
2.7. Ionization Option I
To use the ionization option I unscrew the hose
nozzle and connect the ionization supplement in
its place. Connect the plug to the relevant bush.
Put the oxygen mask on the ionziation head.
We recommend to apply an anti-static spray to
the mask every time before it is used. Additional-
ly an anti-bacteria filter can be inserted.
Do not use a humidifyer during ionization.
The ionization can be activated by pressing I/O
on the remote control supplied. The treatment in-
tervals are set with the buttons marked
and
in 5 minutes steps between 5 and 95 minutes.
The treatment begins by pressing Start. The re-
maining treatment time is indicated in the dis-
play. The treatment can be interrupted by pres-
sing Pause and continued by pressing Start.
An acoutsic signal indicates the end of the tre-
atment and End of Treatment appears in the dis-
play.
The detachable ionization head should be clea-
ned once a week or when used by a different pati-
ent (see also 5. Cleaning).
3. Diagnosis
An extensive autodiagnosis system has been im-
plemented into the oxygen concentrator (PRECI-
SE 6000).
By pressing the button next to the oxygen out-
let on the right the diagnosis is called up.
The displays can be called up by pressing the
button.
Display 1: Service Telephone
Display 2: Software Version,
Equipment Version
Display 3: Date, Time
Display 4: Oxygenconcentration at Option M
Display 5: Pressure Evaluation
Display 6: Working Hours Counter
Display 7: Interval Time
Display 8: 1. Error Memory Bank
Display 9: 2. Error Memory Bank
Display 10: 3. Error Memory Bank
4. Troubleshooting Guide
Symptom
Cause
Remedy
Device does not run
No line voltage
Check if mains cable is
Display: MAINS
connected correctly
Interval Sound
In case check main fuse
Defective fuse
Call service
Device is running
Flow too high
Set to correct flow
Display: 02
82%
Compressor off
Oxygen concentration
Turn device on and off
Display: O2
72%
less 72%
call service
Device is running
Flow regulator turned off
Set to correct flow
Display: 02 Flow too low
Oxygen hose break
Check hose