EasyOne Pro™ / EasyOne Pro™ LAB
97
Hygiene, Cleaning, Maintenance, Disposal
Selecting a Test Subject and Performing the Calibration Check
Start the program.
Select the
Utilities
menu.
Open the
Calibration Test
menu and select
Biological Calibration
Check
.
The
Biological Calibration Check
menu (BioCal-Check main screen) is open.
Select the device according to its serial number
a
(check label on
sensor).
Select the BioCal Test Subject
e
. If the subject is not in the list, press
Add
New BioCal Subject h
and select the subject or add a new one.
Recommendation: Set the Patient ID to "Bio-1", "Bio-2", …
Afterwards return to the BioCal-Check main screen.
Click
Add New Test
d
.
Perform the test (test selection: FVC, FVL, DLCO, Test-End, … according to
configuration settings)
Recommendation: Try to blow a QC grade "A" test.
Recommendation: Perform only one BioCal-Check test a day.
Press
Finish Test
.
To show the graphic trend, click the
Graph
button
g
.
To select another review parameter, change the selection in
Displayed
Parameter Selection
f
.
If you selected a spirometry parameter, the spirometry BioCal quality
grades will be displayed. If you selected a DLCO parameter, the DLCO
BioCal quality grades will be displayed.
The parameter values are divided into two tab sheets
b
:
Baseline Values
(used to calculate warning levels) and
Control Values
(controlled values for
BioCal-Check QC grade calculation).
To deselect a Test (Baseline and Control Values) remove the check mark
c
of
the corresponding test. A deselected test is not used for calculations any
more. This means also that by deselecting a baseline value, the first control
value is moved to the baseline values, because a minimum number (20) of
baseline tests is required for a valid analysis.