Hygiene, Cleaning, Maintenance, Disposal
90
EasyOne Pro™ / EasyOne Pro™ LAB
Maintenance
Before each use
Before each use, visually inspect the device, the cables, the tubing and the
sensor for signs of mechanical damage. Particularly inspect the gas
cylinder(s) and fittings.
If you detect damage or impaired functions that may adversely affect the
safety of the patient or user, do not use
EasyOne Pro™
before it has been
repaired.
Once a year
The following parts must be replaced once a year (see
the gas supply tube
the one-way valve.
Every three years
The filter pack at the back of the device must be replaced every three
years. It consists of the air inlet with filter and a Nafion tube to dry the
sampled gas. Open the screw on the right to replace the filter pack
(Fig. 7-1).
Fig. 7-1
Screw to replace the filter pack