Introduction
Warning
Shock Hazard—
Before using the system, make sure that it is in correct
working order and operating condition. Check the cables
and connectors, in particular, for signs of damage.
Replace damaged cables and connectors immediately,
before use.
Do not expose
EasyOne Pro™
to direct sunlight to
prevent system components from reaching inadmissible,
high temperatures. Furthermore,
EasyOne Pro™
has no
additional protection against the ingress of humidity.
When disconnecting the device from the power line,
remove the plug from the wall outlet first, before
disconnecting the cable from the device. Otherwise there
is a risk of coming in contact with line voltage by
inadvertently introducing metal parts in the sockets of the
power cord.
Do not use multiple portable socket outlets (MPSO) to
connect the devices to the power line.
Printers operated in the patient vicinity must meet the
requirements of IEC 60601. If they do not, they must be
modified and be connected to earth ground (PE, potential
equalization).
All devices of a system must be connected to the same
power supply circuit. Devices that are not connected to
the same circuit must be electrically isolated when
operated, e.g., with an isolated RS232 interface (this is
not a requirement in the USA).
Devices may be connected to other devices or to parts of
systems only when it has been made certain that there is
no danger to the patient, the operators, or the
environment as a result. In those instances where there is
any element of doubt concerning the safety of connected
equipment, the user must contact the manufacturers
concerned or other informed experts to find out whether
there is any possible danger to the patient, the operator,
or the environment as a result of the proposed
combination of equipment. Standards IEC 60601-1-1/
EN60601-1-1 must be complied with in all cases.