Electromagnetic Compatability

8.7

8

Electromagnetic Capability

Troubleshooting & technical information

 The medi pcs requires special precautions regarding EMC and needs to be installed and put into service according to the EMC 

information provided in these accompanying documents.

WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) 

should be used no closer than 30 cm (12 inches) to any part of the medi pcs, including cables specified by the manufacturer.  

Otherwise, degradation of the performance of this equipment could result.

WARNING: A risk of increased emissions or decreased immunity may result if any additional cables are attached.

WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in 

improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are 

operating normally.

WARNING: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this 

equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and 

result in improper operation.

1

2

3

Emissions Test

Compliance

Electromagnetic Environment - Guidance

4

RF emissions

CISPR 11

Group 1

The medi pcs uses RF energy only for its internal function.  There-

fore, its RF emissions are very low and are no likely to cause any 

interference in nearby electronic equipment.

6

RF emissions

CISPR 11

Class B 

7

Harmonic emissions

IEC 61000-3-2

Class A

Voltage Fluctuations 

 IEC 61000-3-3

Complies 

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

The medi pcs is intended for use in the electromagnetic environment specified 

below.  The customer or the user of the medi pcs should assure that it is used in 

such an environment.

The medi pcs is suitable for use in all establishments, including 

domestic establishments and those directly connected to the 

public low-voltage power supply network that supplies buildings 

used for domestic purposes.

Summary of Contents for Pcs brio

Page 1: ...medi pcsbrio medi pcs brio pneumatic compression system Operator s Manual...

Page 2: ...2...

Page 3: ...session 6 4 Turning off the PCU 6 5 Error screen 7 1 Storing the medi pcs brio 7 2 Cleaning the medi pcs brio 8 1 Troubleshooting 8 2 Warranty return policy 8 3 Safety warnings precautions 8 4 Labels...

Page 4: ...stingpulmonary edema existingpulmonaryembolism inflammatoryphlebitis severecongestivecardiacfailure severeatherosclerosisor otherischemicvasculardiseases anycircumstancewhere increasedvenousandlymphat...

Page 5: ...ression garment sizing and application information 4 Power Supply The power supply 2 1 4 connects the PCU to an electrical outlet Step 1 When your medi pcs brio arrives carefully unpack the contents a...

Page 6: ...he garment going up the garment from distal to proximal The step value setting can be changed in 1 mmHg increments ranging from 1 60 mmHg However the step value range is limited such that no chamber w...

Page 7: ...distal pressure is set to 50 mmHg with a step value of 3 mmHg Chamber 1 2 3 4 5 6 mmHg 50 47 44 41 38 35 Example 2 An 8 chamber garment whole leg or arm where the distal pressure is set to 80 mmHg wit...

Page 8: ...e medi pcs brio is now ready to operate Please review the settings displayed in the Treatment Summary screen to ensure that all settings have been entered correctly If you must unlock or change the tr...

Page 9: ..._____________________________ Language ______________________________ Treatment time ____________________________ Distal pressure ______________________________ Access level __________________________...

Page 10: ...cm Lower Leg Sizing Measurements T i p o f M i d d l e F i n g e r t o A x i l l a C G Floor to Knee Crease B C Floor to Gluteal Fold C G B Arm Garment Sizing Measurements Size Item Leg Length Circumf...

Page 11: ...before applying If desired apply a liner over the limb before applying the garment This will help keep the garment clean during use Loose unrestrictive clothing such as sweatpants or cotton t shirts m...

Page 12: ...arment either slide your limb out of the garment or undo the hook and loop material and remove your limb It is recommended to keep the hook and loop material fastened even when the compression garment...

Page 13: ...e latch mechanism and pulling the connector out The blocking port can remain in place for future treatment sessions Step 1 Locate the garment connectors at the end of the hoses attached to the garment...

Page 14: ...appear 6 4 The PCU will begin the fill to fit process after which the programmed treatment cycle will begin The Treatment Running screen will show treatment time remaining current active chamber in us...

Page 15: ...t Summary Screen Press the power switch on the back of the PCU to turn off the medi pcs brio Please note that the PCU inflation deflation valves close while the PCU is turned off If a garment is conne...

Page 16: ...sure that the power supply is unplugged before cleaning the PCU Gently wipe the PCU with a slightly damp cloth Do not allow any moisture or liquid to enter the device If additional cleaning is require...

Page 17: ...d any rapid movements or drastic posture changes during the first 5 10 minutes of treatment 2 Verify that the correct treatment settings have been selected by referring to the Treatment Summary screen...

Page 18: ...e or service the device while in use To completely eliminate power disconnect the power supply from the Pressure Control Unit or from the wall outlet Warning Electric Shock Hazard Unplug power supply...

Page 19: ...nce pain or notice your condition worsening contact your physician Do not change your treatment settings without consulting your physician Warning Parental Supervision Children under the age of 18 sho...

Page 20: ...ing the safety requirements specified for Class II equipment according to IEC 61140 Non ionizing electro magnetic radiation ISO 7000 IEC 60417 Graphical symbols for use on equipment 5140 To indicate o...

Page 21: ...47 Part15 Subpart B ICES 001 ISSUE 4 AS NZS CISPR 11 JIS T 0601 1 2 ABNT NBR IEC 60601 1 2 LEGAL MANUFACTURER medi USA LP 6481 Franz Warner Parkway Whitsett NC 27377 www mediusa com Made in USA Tempe...

Page 22: ...perating normally WARNING Use of accessories transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or...

Page 23: ...cle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles h Single phase at 0 0 UT 250 300 cycle 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles...

Page 24: ...ncy range b Interference may occur in the vicinity of equipment marked with the following symbol Guidance and Manufacturer s Declaration Electromagnetic Immunity NOTE 1 At 80 MHz and 800 MHz the highe...

Page 25: ...o any part of the medi pcs including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance E 6 d P...

Page 26: ...nd mobile RF communications equipment transmitters and the medi pcs as recommended below according to the maximum output power of the communications equipment For transmitters rated at a maximum outpu...

Page 27: ...27 For other language translations please visit www mediusa com...

Page 28: ...medi USA 6481 Franz Warner Parkway Whitsett NC 27377 info mediusa com T 800 633 6334 F 888 570 4554 mediusa com PBO157 Rev A...

Reviews: