Batteries
The battery indicator shows how much power is available to the
wheelchair.
• Red, amber and green indicate that the batteries are fully charged.
• Red and amber indicate that the batteries will soon need recharging.
• Red indicates that the batteries need to be recharged as soon as
possible, otherwise the wheelchair will cut out.
The MC Concept’s batteries should be charged while not in use.
This extends the service life of the batteries.
The MC Concept is supplied with sealed, maintenance-free
batteries that do not normally generate gas and do not need to
be topped up with water.
Note!
Unsealed batteries must NEVER be installed in the MC Concept.
If charging is to take place outdoors, an enclosed charger
without a fan should be chosen.
Only use chargers designed for charging dry maintenance-free
GEL batteries.
Max. charging current 12 A.
The charging cable must NOT be extended.
New batteries can be purchased from Medema Danmark A/S.
Battery disposal
Used batteries must be disposed of through your supplier or at
a recycling centre.
Take care in handling any leaky batteries, as they contain
corrosive acid bound into the gel mass.
INFO: New batteries can be purchased from Medema Danmark
A/S.