User guide P9-0222-B
20 of 64
Version 1.0.9/2015
Medema A/S
Yearly
The MC Concept 1122 is designed to stay safe to use for at
least 10 years, up to a maximum of 5,000 hours, providing it has
service and safety inspections every year, which is equivalent
to around 500 operating hours. The service must be carried out
either by Medema Danmark A/S or an authorised workshop.
IMPORTANT! For safety reasons it is of the utmost importance
that service and safety check intervals are complied with, as this
minimises the risk of brake failure and short-circuits in the wiring,
which could generate heat and cause a fire.
Programming
The Dynamic electronic system is programmable so that the
driving characteristics can be optimised for individual users.
Programming must only be carried out by people trained in
Dynamic’s control systems. Changing parameters incorrectly
can produce driving characteristics that are dangerous for the
user.
Note
Mobile phones should not be used very close to the wheelchair
as they generate electromagnetic fields that may affect the
electronic system.
The electronic system of the wheelchair itself also generates
some electromagnetic fields, which on rare occasions may
affect equipment like the alarm systems in shops.
Dynamic’s control systems have been tested and meet the
requirements of ISO7176/14 and EN12184