15
Vibrogram Measurement completion
When the measurement is finished, click on Vibrogram
Settings and Close the Test Settings window. The
following window will appear (figure 13), select Yes. It
is not intended to leave the test settings in the Amadé.
As the Vibrogram thresholds have been received as a
basis for the first fitting it can now be continued with
entering the thresholds (and UCL values) in the
Audiogram tab.
figure 13
The Audiogram Tab
Select the Audiogram; this is the first tab in the Connexx
software and it will appear after creating a new client.
The data entered in this screen will be used by the
Connexx software to calculate the target gains for the
Amadé. For the first fitting the thresholds of the patient
have to be entered at least on the side which is to be
fitted (figure 14).
figure 14
When fitting the Amadé audio processor of the VIBRANT
SOUNDBRIDGE the thresholds and the dynamic range
of the patient (UCL) are defined by the Vibrogram. It is
therefore suggested to use the Vibrogram as a basis for
the fitting as it includes the hearing thresholds of the
patient as well as the efficiency of energy transfer from
the FMT to the inner ear.
Air Conduction symbols are used to enter the Vibrogram
thresholds, if UCL values have been measured, UCL
symbols are used to enter them here.