MEDAP 5752 4916 Operating Instructions Manual Download Page 32

Manufacturer:

ATMOS 

MedizinTechnik GmbH & Co. KG 

Ludwig-Kegel-Str. 16 

79853 Lenzkirch 

GERMANY 

Phone: +49 7653 689-0 

www.atmosmed.com

Summary of Contents for 5752 4916

Page 1: ...OPERATING INSTRUCTIONS MEDAP LS FLOW FLOWMETER GA 5752 4981 GB 15 ...

Page 2: ... 4981 GB 15 2 Subject to technical modification Illustrations and technical specifications may vary slightly from those in these Operating Instructions as a result of ongoing product development V15 2019 12 4 ...

Page 3: ...8 1 5 1 Use in accordance with the intended purpose 8 1 5 2 Applicable standards 9 1 5 3 Intended purpose 9 1 5 3 1 Possible applications 10 1 5 4 Versions of LS FLOW LS DFLOW flowmeter 10 1 5 5 Interface description 11 1 5 5 1 Approved interface for compressed gas supply 11 1 5 5 2 Flowmeter outlet 11 1 5 5 3 Connection tube 12 1 5 5 4 Tube adapter for Air and O2 12 2 Safety notes 13 2 1 General ...

Page 4: ...ning and disinfection 21 6 1 General 21 6 2 Cleaning 22 6 2 1 General 22 6 2 2 Cleaning procedure 22 6 3 Disinfection 22 6 3 1 General 22 6 3 2 Suitable disinfectants 23 6 3 3 Disinfection procedure 23 6 4 Special safety notes 24 7 Maintenance 25 7 1 General 25 7 2 Period tests 25 7 3 Malfunctions and troubleshooting 25 7 4 Repairs 26 7 5 Service hotline 26 7 6 Type plate position 26 7 7 Sending i...

Page 5: ...ions carefully and completely before using the product for the first time Always proceed in accordance with the information contained herein Store these operating instructions in a location near the product 1 2 1 Abbreviations EN European standard EEC European Economic Community PD Punched disc 1 2 2 Symbols 1 2 2 1 Cross references References to other pages in these operating instructions are ide...

Page 6: ... Structure of notes Notes not referring to personal injury or property damage are structured as follows Pictogram Descriptor Reference to NOTE Supplementary assistance or further useful information ENVIRONMENT Information regarding proper disposal Tab 2 Structure of notes 1 2 4 Symbols used Symbols are attached to products type plates and packaging Symbols Identification 4 Labelling for products w...

Page 7: ...ance with the ISO 15223 1 standard Symbol for Relative humidity Labelling in compliance with the ISO 15223 1 standard Symbol for Atmospheric pressure Meaning of type plate Symbol for Oil and fat free Labelling in compliance with the ISO 15223 1 standard Symbol for Product number Labelling in compliance with the ISO 15223 1 standard Symbol for Name and address of the manufacturer as well as date of...

Page 8: ...dow 3 Gas probe 4 Flowmeter outlet Version B Tapping unit with rail clamp and NIST connection 5 Equipment rail 6 NIST connection 7 NIST nipple 8 NIST screw connection 9 Connection tube 10 Tube adapter 4 mm 6 mm 8 mm REF 5752 2746 11 Tube adapter plastic 4 mm 6 mm 8 mm REF 5752 5316 Version C Tapping unit with NIST connection for screwing without rail clamp 1 5 Basic requirements 1 5 1 Use in accor...

Page 9: ...sols Medical indications application Inhalation and insufflation of oxygen within the scope of oxygen enrichment via an inhalation mask or a nose latch for patients breathing by themselves Together with a hand held nebuliser administration of water soluble drugs via an inhalation mask Specification of the main function Supply of oxygen or compressed air For oxygen compressed air supply the LS FLOW...

Page 10: ...ate tubes with inner diameters of 4 mm 6 mm and 8 mm may be connected to the flowmeter via a tube adapter metal or plastic With the bubble humidifier REF 5752 5315 for humidifying oxygen The bubble humidifier is intended for inhalation The bubble humidifier is connected without an adapter Sterile water systems disposable humidifiers UNF 9 16 may be connected directly 1 5 4 Versions of LS FLOW LS D...

Page 11: ... 5752 5303 LS FLOW A 15 NIST connection screw connection REF 5752 5709 LS DFLOW A 15 Wall DIN REF 5752 5607 LS DFLOW A 15 NIST connection equipment rail REF 5752 5608 1 5 5 Interface description All devices and accessories which are combined with the tapping unit must be listed in the accessories list or meet the specifications of the interface description The configuration of the overall system a...

Page 12: ...r must be pressure resistant and must comply with the hygiene standard of the hospital The inside diameter of the connection tube must match the outside diameter of the tube adapter 1 5 5 4 Tube adapter for Air and O2 The tube adapter is used to connect the tapping unit and the connection tube The inner thread of the tube adapter must match the outer thread of the tube connector of the flowmeter o...

Page 13: ...s react explosively with oils greases and lubricants Fire and explosion hazard due to compressed gases Always keep the product free of oils greases and lubricants DANGER Risk of fire Escaping oxygen increases the risk of fire Never smoke near equipment which carries oxygen and avoid using open fires or glowing objects Check tight fitting and firm seat of the connector when mounting accessories DAN...

Page 14: ...bility and tightness of the product cannot be guaranteed if the maximum upper and lower ambient temperatures are exceeded WARNING Impacts Impacts may cause damage to sensitive precision mechanical components Do not expose the product to impacts WARNING Non permissible load If the permissible load is exceeded leakages may occur at the connection between terminal unit and gas probe In accordance wit...

Page 15: ...heck whether all tubes are undamaged _ Check whether the unit has been properly cleaned and that there are no residues or soiling _ Do not use damaged components 3 2 Connection to the terminal unit 3 2 1 General NOTE Please refer to the manufacturer s instructions for the particular terminal unit for information on connecting the gas probe to the terminal unit 3 2 2 Version A 1 1 2 2 Fig 2 Version...

Page 16: ...pping unit is in a stable position on the equipment rail _ Insert the NIST nipple 4 of the connection tube into the NIST connection 5 of the tapping unit and tighten down the cap nut 6 by hand _ Plug the gas probe 7 into the terminal unit 8 3 2 4 Version C Fig 4 Version C Tapping unit with NIST connection for screwing _ Insert the NIST nipple 1 of the connection tube into the NIST connection 2 of ...

Page 17: ...difier contains a flat seal _ Fill the humidifier bottle up to the marking Filling level with distilled water and screw the bubble humidifier cap into place _ Screw cap nut 1 of the bubble humidifier to the flowmeter output 2 and tighten finger tight until stop _ Pay attention that no tensile forces affect the bubble humidifier 3 3 3 Connection of disposable humidifiers from other manufacturers Di...

Page 18: ...tightly fastened The tapping unit is locked firmly to the equipment rail Version C Tapping unit with NIST connection for screwing The gas probe of the connection tube is correctly inserted into the terminal unit The NIST nipple of the connection tube is correctly secured in the NIST connection of the tapping unit The NIST screw connection is tightly fastened The tapping unit is screwed tight All v...

Page 19: ... WARNING Compressed gas setting The regulating mechanism is sensitive Make the compressed gas settings very carefully NOTE The flow rate can be set to the following 14 values 0 0 2 0 5 0 7 1 1 5 2 3 4 5 7 10 12 15 at 15 l min 0 0 1 0 2 0 3 0 4 0 6 0 8 1 1 5 2 2 5 3 4 5 at 5 l min 0 0 1 0 15 0 2 0 25 0 3 0 35 0 4 0 5 0 6 0 7 0 8 0 9 1 at 1 l min NOTE When applying oxygen in its function as a medica...

Page 20: ... protect the product from damages cover with cloths when storing NOTE Refer to the medical gas distributor instructions for information regarding detaching the gas probe from the terminal unit All versions _ After having completed the treatment close the adjustment screw by turning it clockwise and check to ensure that it is closed _ Disconnect the tapping unit gas probe from the terminal unit Ver...

Page 21: ...gas Residues of sterilisation gases or other substances in the unit could endanger the patient s health Do not disassemble the product and do not clean it by gas sterilisation or any other mechanical cleaning or sterilisation method Ensure that no disinfectant or cleaning agent enters the product CAUTION Improper cleaning and disinfection can cause property damage Do not use the following products...

Page 22: ...ning can cause property damage Do not spray cleaning agent directly into the joints or gaps and never use a high pressure cleaning unit 6 2 2 Cleaning procedure _ Use the correct dose of all purpose detergent with water for the degree of surface contamination and in accordance with the instructions of the detergent manufacturer _ Thoroughly wipe off the product with a soft cloth slightly wetted in...

Page 23: ...uaternary compounds Alkyl didecyl polyoxethyl ammonium propionate alkyl dimethyl alkylbenzyl ammonium chloride alkyl dimethyl ethyl ammonium chloride alkyl dimethyl ethylbenzyl ammonium chloride benzalkonium propionate benzalkonium chloride alkyl dimethyl benzyl ammonium chloride coco dimethyl benzyl ammonium chloride lauryl dimethylbenzyl ammonium chloride myristyl dimethyl benzyl ammonium chlori...

Page 24: ...aterial damage Various components in the product are made of plastic materials Solvents and some disinfectants and some cleaning agents can soften plastic or cause tension fissures Never use detergents that contain alcohol Observe cleaning and disinfection specifications CAUTION Property damage Using non colour fast surgical drapes can cause discolouration of surfaces Only use colour fast surgical...

Page 25: ...oting Defect Source of malfunction Troubleshooting No flow Flow rate is set to the value 0 Set the flow rate to the desired value Product defective Have the equipment repaired by a service technician authorised by ATMOS Flow does not stop Flow rate is not set to the value 0 Set the flow rate to the value 0 Product defective Have the equipment repaired by a service technician authorised by ATMOS Th...

Page 26: ...apter Sending in the device page 26 7 5 Service hotline 49 7653 689 0 7 6 Type plate position 1 Fig 7 Type plate Position of the type plate 1 7 7 Sending in the device _ Remove and properly dispose of consumables _ Clean and disinfect the product and accessories according to the operating instructions _ Place used accessories with the product _ Fill in the form QD 434 Delivery complaint return shi...

Page 27: ...ent MR safe to a limited extent tapping unit without humidifier adapter accessories up to 4 7 T 100 kPa 1 bar 1000 mbar 750 mmHg 1T 1000mT 10000 Gauss 8 3 Ambient conditions Temperature 40 C to 60 C shipping storage 20 C to 60 C operation Relative humidity less than 100 shipping storage 30 to 75 operation Atmospheric pressure 700 hPa to 1060 hPa shipping storage 700 hPa to 1060 hPa operation 8 4 D...

Page 28: ...Weight 220 g LS FLOW flowmeter version B P5 version Dimensions 40 x 100 x 120 mm Weight 306 g LS FLOW twin flowmeter version A Dimensions 125 x 65 x 100 mm Weight 540 g LS FLOW twin flowmeter version A P5 version Dimensions 125 x 56 x 100 mm Weight 512 g LS FLOW twin flowmeter version B Dimensions 125 x 100 x 140 mm Weight 690 g ...

Page 29: ...ories The following accessories are not part of the scope of delivery and must be ordered separately 9 1 Accessories 5752 2746 Tube adapter metal 4 mm 6 mm 8 mm 5752 5316 Tube adapter plastic 4 mm 6 mm 8 mm 5752 5315 Bubble humidifier 5752 5707 Flow selector Tab 6 Accessories ...

Page 30: ...Notes ...

Page 31: ...Notes ...

Page 32: ...Manufacturer ATMOS MedizinTechnik GmbH Co KG Ludwig Kegel Str 16 79853 Lenzkirch GERMANY Phone 49 7653 689 0 www atmosmed com ...