Cleaning and disinfection
General
GA 5752 4981 GB 15
21
6
6
Cleaning and disinfection
6.1
General
The product must be wipe or spray disinfected after every use.
DANGER!
Risk due to incorrect use of detergents and disinfectants!
It is strictly advised to observe the manufacturer instructions regarding how to use
the detergents and disinfectants as well as the valid hospital hygiene rules.
DANGER!
Infection hazard!
Product may be contaminated.
Always wear gloves for cleaning and disinfection.
DANGER!
Infection hazard!
Particles of grime may become encapsulated and lead to the product not reaching
the desired germ-reduction after disinfection.
Before disinfection, the product must be cleaned thoroughly of contamination and
encapsulated particles of grime.
DANGER!
Health hazard!
The product is also used to administer respiratory gas. Residues of sterilisation
gases or other substances in the unit could endanger the patient's health.
Do not disassemble the product and do not clean it by gas sterilisation or any other
mechanical cleaning or sterilisation method. Ensure that no disinfectant or cleaning
agent enters the product.
CAUTION!
Improper cleaning and disinfection can cause property damage!
Do
not
use the following products for cleaning and disinfection:
• Products containing alcohol (e.g. hand disinfectants)
• Halogenides (e.g. fluorides, chlorides, bromides, iodides)
• Dehalogenating compounds (e.g. fluorine, chlorine, bromine, iodine)
• Products that may scratch the surface (e.g. scouring agents, wire brushes, wire
wool)
• Standard commercial solvents (e.g. benzene, thinner)
• Water containing iron particles
• Cleaning sponges containing iron
• Products containing hydrochloric acid
Use a soft, lint-free cloth or a soft nylon brush to clean the product.
CAUTION!
Improper cleaning and disinfection can cause property damage!
Use only as much detergent and disinfectant as required.