Taking the unit out of operation
Taking the unit out of operation
20
GA 5752 4981 GB 15
5
5
Taking the unit out of operation
5.1
Taking the unit out of operation
WARNING!
Disengage!
When the product is disengaged (removed) from the terminal unit, the pressure
energy may cause recoil.
Use the terminal unit in parking position or support when disengaging.
NOTE
To protect the product from damages, cover with cloths when storing.
NOTE
Refer to the medical gas distributor instructions for information regarding detaching
the gas probe from the terminal unit.
All versions:
After having completed the treatment, close the adjustment screw by turning it clockwise and
check to ensure that it is closed.
Disconnect the tapping unit / gas probe from the terminal unit.
Version B: Tapping unit with rail clamp
Remove the unit from the equipment rail. For this purpose, open the handle screw and lift the
unit off the equipment rail.