Operation
Setting the flow for treatment
GA 5752 4981 GB 15
19
4
WARNING!
Risk of injury!
Accessories of the product (e.g. humidifier, adapter, hand-held nebuliser,
connection tubes, accessories) may be affected by the magnetic field.
If the product is used in conjunction with accessories within the 0.5 mT line, all
connected accessories must be MR compatible. Observe the operating instructions
of all connected accessories or consult the manufacturer of the product.
NOTE
If the product is used in conjunction with accessories outside the 100 mT line, the
product does not create artefacts on the MR images.
4.3
Setting the flow for treatment
WARNING!
Compressed gas setting!
The regulating mechanism is sensitive. Make the compressed gas settings very
carefully!
NOTE
The flow rate can be set to the following 14 values:
0; 0.2; 0.5; 0.7; 1; 1.5; 2; 3; 4; 5; 7; 10; 12; 15 (at 15 l/min),
0; 0.1; 0.2; 0.3; 0.4; 0.6; 0.8; 1; 1.5; 2; 2.5; 3; 4; 5 (at 5 l/min),
0; 0.1; 0.15; 0.2; 0.25; 0.3; 0.35; 0.4; 0.5; 0.6; 0.7; 0.8; 0.9; 1 (at 1 l/min),
NOTE
When applying oxygen in its function as a medication, as per the monograph in the
European Pharmaceuticals Reference it is absolutely necessary to measure the
flow rate.
1
Fig. 6: Working with the unit
Use the rotation regulator to set the flow rate
to the required value for the treatment.
Increasing the flow rate
Turn the rotation regulator (1) anti-
clockwise.
The flow rate is increased.
Reducing the flow rate
Turn the rotation regulator (1) clockwise.
The flow rate is decreased.